User Guide

2524
To clear the numeric data
entry, press CLEAR key.
The instrument should be
calibrated weekly or every
time the probe has been
changed.
To quit from calibration
mode, press the CAL key.
PROCEDURE WITHOUT ATC (MTC)
If you need to calibrate without ATC,
press the ATC/MTC key. The decimal
point "." of the reading will blink on the
TEMP. display (both in the Celsius and
in the Fahrenheit scale).
During calibration, once the selected
buffer value is confirmed, "25°C" or
"77.0°F" will start blinking on the TEMP.
display to prompt the selection of the
calibration temperature value.
Take the temperature of
the solution with a
Checktemp or a glass
thermometer.
Enter the correct buffer
temperature value by
using the numeric keys
and press ENTER to
confirm the temperature
setting.
Continue the procedure as for calibra-
tion with ATC but without temperature
probe connection.
PGM
Clear
CAL
ATC
MTC
0
ENTER
The conductivity calibration procedure is now
complete.
Notes:
If the buffer dif-
fers by more than
±15% from the
set value, the "
BUF
"
symbol will blink
on the display.
Press the CAL
key to exit the
calibration mode
and repeat the
procedure using
the correct buffer.
If the HI7669/2W temperature probe
is not connected to the meter, once
confirmed the selected buffer value,
"25°C" or "77.0°F" will start blinking on
the TEMP. display to prompt the se-
lection of the calibration temperature
value.
Take the temperature of
the solution with a
Checktemp or a glass
thermometer.
Enter the correct buffer
temperature value by
using the numeric keys
and press ENTER to
confirm the temperature
setting.
The calibration procedure is complete.
ENTER
CAL
CONDUCTIVITY
%TC
S
m
CAL
BUF