Method ID: HI8001EN 
5.0 mg/mL Titrant Standardization using a Liquid Water Standard 
for one-component titrant 
  Titrator Application Methods: General Methods - r. 0.4 
Description: 
Method  for  the  standardization  of  5.0  mg/mL  One-
Component Karl Fischer Titrant using a Liquid Water 
Standard. The results are expressed in mg/mL. 
Electrode: 
  HI 76320 Double Platinum Pin Electrode 
Reagents: 
  5 mg/mL One-Component KF Volumetric Titrant 
  10 mg/g Liquid Water Standard 
  Dry Methanol 
Other Accessories: 
  A clean, dry 3-mL syringe 
  A  clean,  dry  22-gauge,  6”  non-coring  septum 
penetration needle  
  GL 45-Thread Solvent Bottle 
Procedure: 
  Setup titrator according to the instruction manual. 
  Press “Select Method” from the main screen. Use 
the  arrow  keys  to  highlight  ‘HI8001EN  Stdz 
5mg/mL w/ water Std.’ and press “Select”. 
  Install  a 5-mL burette filled with  5  mg/mL One-
Component  KF  Volumetric  Titrant  to  be 
standardized and verify that no air bubbles are 
present in the burette or tubing. If necessary prime 
until all air has been removed completely. 
  Connect the solvent bottle top assembly to the 
bottle of methanol according to the manual. 
  Prepare  the  titration  vessel  according  to  the 
manual. Dispense enough solvent from the solvent 
bottle to  fill  the  vessel  to the  “min” line (about 50 
mL). 
  Press “Start/Stop” to pre-titrate the solvent and the 
titration vessel  moisture.   Allow the background 
drift rate to stabilize before proceeding to the next 
step. 
  Fill  the  syringe  and  needle  with  the  water 
standard. 
  Weigh the syringe, needle and water standard. 
  Press  “Start  Analysis”.    You  will  be  prompted  to 
enter the sample size. 
  Dispense 1.00 g (about 1 mL) of standard into the 
titration  vessel  through  the  septum  using  the 
needle.    Pay  attention  not  to  get  any  sample  on 
the electrode  or  beaker  walls.    If  necessary swirl 
the titration vessel by hand. 
  Clear the needle of residual standard by intaking a 
small volume of  air from the titration vessel.  If a 
“hanging  drop”  of  standard is seen on the end of 
the needle, dip the end of the needle briefly in the 
solvent. 
  Remove  the  needle  from  the  titration  vessel  and 
weigh the syringe again in order to determine the 
added standard mass (by difference of the two 
measurements.) 
  Use the numeric keypad to enter the exact weight 
and press “Enter” to start the analysis. 
  At  the end  of  the titration, the  “Standardization 
Result”  screen  is  displayed.    The  results  are 
expressed in mg/mL. 
Method Parameters:  
Name:  Stdz 5mg/mL w/ water Std 
Method Revision:  1.0 
Type:  Titrant Standardization 
Predispensing Amount:  25 % 
Pre-Analysis Stir Time:  5 Sec 
Stirring Speed:  900 RPM 
Stirbar Type:  Medium 
Drift Entry:  Automatic 
Solvent:  Methanol 
Standard:  Liquid 10 mg/g 
  Type:  Liquid by mass 
  Concentration Unit:  mg/g 
  Water Content:  10.0000 mg/g 
Standard Size:  1.0000 g 
Titrant:  Composite 5 
  Titrant Type:  One Component 
  Nominal Titrant Conc.:  5.0000 mg/mL 
  Std. Titrant Conc.:  5.0000 mg/mL 
  Date/Time:  Jan 1, 2011 12:00 
  Titrant Age Reminder:  2d:00h:00m 
Control Parameters: 
  Start Mode:  Normal 
  Standby Mode:  Enabled 
  Standby Duration:  720 minutes 
  Imposed Current:  20 uA 
  Minimum Dose:  0.500 uL 
  Maximum Dose:  20.000 uL 
  Timed Increment:  1 second 
  End Point Value:  180.0 mV 
  Signal Averaging:  3 Readings 
  Flow Rate:  10.0 mL/min 
Termination Parameters: 
  Maximum Duration:  1200 sec 
  Maximum Titrant Volume:  10.0000 mL 
  Term. Criterion:  Relative Drift 
  Relative Drift:  7.0 ug/min 
Calculations: 
Titrant units:  mg/mL 
Titrant volume consumed:  V (mL) 
Final Results Units:  mg/mL 
Standard Concentration:  10.0000 mg/g 
Standard mass:  1.0000 g 