User Guide
39
• Adjust the OFFSET trimmer if the LCD does not show the zero
reading.
• Pour enough conductivity solution into a
beaker to achieve at least 8 cm (3¼") of
depth.
• Immerse the probe into the beaker with the
conductivity solution. The holes on the sleeve
must be completely submerged in the solution.
• Tap the probe repeatedly on the bottom of the beaker and stir it
to ensure that no air bubbles are trapped inside the sleeve.
• If the temperature of the probe is close to that of the solution the
display will stabilize quickly and provide you with a temperature
compensated conductivity measurement.
However, allow a few minutes if there is a temperature difference
of 5°C (9°F) or more for the ATC circuitry to compensate completely.
• When the reading stabilizes, turn the SLOPE trimmer on the front
of the HI943500 until the LCD reading is the same of the
calibration solution at 25°C (77°F), i.e.:
80.0 mS using HI7034 with HI943500A
12.88 mS using HI7030 with HI943500B
1413 µS using HI7031 with HI943500C
84.0 µS using HI7033 with HI943500D
8 cm
(3¼
")
HI 7030