• Pour enough conductivity solution into a
plastic beaker to achieve at least 8 cm
(3¼") of depth.
• Immerse the probe into the beaker with the
conductivity solution. The holes on the sleeve
must be completely submerged in the solution.
• Tap the probe repeatedly on the bottom of the beaker and stir it
to ensure that no air bubbles are trapped inside the sleeve.
• If the temperature of the probe is close to that of the solution the
display of HI 8931 will stabilize quickly and provide a temperature
compensated conductivity measurement. However, allow a few
minutes if there is a temperature difference of 5°C (9°F) or more for
the ATC circuitry to compensate completely.
• For HI 8936
Adjust the SLOPE trimmer of the HI8936 transmitter to read on
the ammeter:
"10.4 mA" using HI7034 and HI8936A/AL
"14.304 mA" using HI7030 and HI8936B/BL
"15.304 mA" using HI7031 and HI8936C/CL
"10.72 mA" using HI7033 and HI8936D/DL
8 cm
(3¼
")
HI 7030