User Guide

35
• Leave the HI7638 conductivity probe
in air (dry probe).
• When the power is on, the ammeter should read "4.0 mA" or the
HI8936L transmitter with LCD should display "0".
If not adjust the OFFSET trimmer
of the HI8936 transmitter to ob-
tain "4 mA" or "0" on the HI8936L.
The HI 8931 controller should dis-
play "0" value.
• If not adjust the OFFSET trimmer of
HI8931 to display a zero reading.