CALIBRATION PROCEDURE OF
HI8931 & HI8936 WITH HI7638
Material needed
HI7638 conductivity probe
HI 8931 conductivity controller
HI8936 conductivity transmitter
An ammeter (for transmitters without LCD)
• Calibration solutions, according to the different models:
HI7034 80 mS/cm @25°C
for HI8931A, HI8936A & AL
HI7030 12.88 mS/cm @25°C
for HI8931B, HI8936B & BL
HI7031 1413 µS/cm @25°C
for HI8931C, HI8936C & CL
HI7033 84 µS/cm @25°C
for HI8931D, HI8936D & DL
PROCEDURE
Connect the HI 7638 to the HI8936 transmitter (see page 23).
Connect the HI8936 transmitter to the HI8931 controller (see
page 24).
• Connect the HI8931 controller to the mains (see page 24).
Before proceeding with the calibration,
make sure that the meter is in measure-
ment mode (MEASURE LED will be lit)
and not in set mode.
Connect the ammeter to the HI8936 transmitter to monitor the
signal current (see the following picture).
MEA
SURE