The reading will be converted to mA by the following formula:
mA = K (measured value x 16/2000) + 4
K = conversion factor depending on the model
Model Conversion factor K
HI 8936A & AL 10
HI 8936B & BL 100
HI8936C & CL 1
HI8936D & DL 10
For example, using a HI8936A, if the measured value is 123.4
mS, then
output current = 10 x (123.4 x 16/2000) + 4
= 13.9 mA
Adjust the SLOPE trimmer of the
HI8936 transmitter to read "13.9
mA" on the ammeter or the reading
displayed by the HI8733 conduc-
tivity meter on the HI8936L dis-
play. E.g. "123.4 mS".
Adjust the SLOPE trimmer of the HI 8931 controller until the
controller reading is the same as the reading displayed by the
HI8733 conductivity meter. E.g. "123.4 mS".