User Guide

31
Ensure that there is no solution inside the HI7635 conductivity
probe (dry probe).
When the power is on, the ammeter should read "4.0 mA". The
HI8936L transmitter with LCD should display "0".
• If not, adjust the OFFSET trimmer
of the HI8936 transmitter to ob-
tain "4 mA" or "0" on the
HI8936L.
The HI 8931 controller should dis-
play "0" value.
• If not, adjust the OFFSET trimmer
of HI8931 to display a zero read-
ing.
• Switch the flow on and allow the
sample solution to flow through
the HI7635 conductivity probe.
Collect a sample of this solution in
a beaker.
Measure the conductivity of the
solution with the HI8733 por-
table conductivity meter with ATC.
The value obtained will be used
for the transmitter and controller
calibration.