User Guide
30
CALIBRATION PROCEDURE OF
HI8931 & HI8936 WITH HI7635
Material needed
• HI7635 conductivity probe
• HI 8931 conductivity controller
• HI8936 conductivity transmitter
• A 20 mA f.s. ammeter (for transmitters without LCD)
• A reference conductivity meter with automatic temperature com-
pensation accurately calibrated (e.g. HI8733).
PROCEDURE
• Connect the HI7635 probe to the HI8936 transmitter (see page
23).
• Connect the HI8936 transmitter to the HI8931 controller (see
page 24).
• Connect the HI8931 controller to the mains (see page 24).
Before proceeding with the calibration,
make sure that the meter is in measure-
ment mode (MEASURE LED will be lit)
and not in set mode.
• Connect the ammeter to the HI8936 transmitter to monitor the
signal current (see the following picture).
MEA
SURE