User Guide

26
• It is recommended that the HI 7635 is installed vertically. This is
to ensure that trapped air bubbles or turbulent flows cause
minimal interference to the measurement system. The maximum
working pressure of this unit is 5 BAR (72.5 psi).
CAUTION: do not use in systems where temperature exceeds 80°C
(176°F).
The HI7638 process conductivity probe is also
supplied with a 3m (10') length cable. The
six core cable from this probe is connected to
the HI 8936 process conductivity transmitter
as shown.
The 7-pin DIN connector has to be removed when this probe is
used in conjunction with HI8936 transmitter. See page 23 for the
proper connection scheme.
FOR HI943500 ONLY
Ensure that the HI7638 con-
ductivity probe is connected to
the meter securely by aligning
the pins with the socket, push-
ing the plug in and tightening
the threaded ring.