User Guide

25
The HI 8936 transmitter may be wall-
mounted at any convenient location close
to the measurement site. To minimize
thermal drifts due to extreme temperature
fluctuations, particularly if the measure-
ment is conducted outdoors, it is recom-
mended that the transmitter is protected
in a casing.
For the installation of the HI 7635 conductivity probe, it is
necessary to use rubber seals between the probe and the pipe or
elbow joints. A pipe sealant is also recommended to ensure a leak
free joint. When screwing on the joints, take care not to over-
tighten as excessive pressures can cause the probe to be damaged.
The conductivity probe HI7635 is sup-
plied with a 3m (10') length of cable.
The six core cable from the probe is
connected to the HI8936 process con-
ductivity transmitter as shown. The
connections are color coded for an easy
installation. See page 23 for the proper
connection scheme.