User Guide
21
REAR CONNECTIONS FOR HI 943500
• Power Connection Terminals
4-screw-terminal-strip for connec-
tion to a 3-wire power cable ac-
cording to the indicated voltage
(115 or 230V).
• 7-pin DIN connector
socket
For connection to the
HI 7638 conductivity
probe.
• + mA output -
The first and the second
terminals are the output
terminals for connection to
a recorder or other control
equipment. The output cur-
rent varies from 4 to 20
mA and is proportional to
the measured conductivity
value.
• Set Select
These contacts permit the activation of the Set Contact relay when
the measured value is lower (connected terminals) or higher (open
terminals) than the user's set value. See also page 27.