14
0.0 to 199.9 mS/cm
0.00 to 19.99 mS/cm
0 to 1999 µS/cm
0.0 to 199.9 µS/cm
±2% of Full Scale
excluding probe error
±2.5% of Full Scale
±0.4 mA
HI8936 A or AL (not included)
HI8936 B or BL (not included)
HI8936 C or CL (not included)
HI8936 D or DL (not included)
------
HI7635 for in-line applications or
HI7638 for tank (not included)
Manual 2-point through offset and
slope trimmers
See transmitter HI8936
Automatic from 0 to 60°C with ß=2%
4-digit LCD plus graphic symbols
4 to 20 mA (isolated)
Isolated, 2A, Max. 240V, resistive
load, 1,000,000 strokes
110/115V or 220/240V
(user selectable); 50/60 Hz
141 x 69 mm (5.6 x 2.7")
-10 to 50°C (14 to 122°F)
1 Kg (2.2 lb.)
DIN 43 700 144x72 mm (5.7x2.8") in
black anodized aluminum. Front and
back with shockproof ABS plastic and a
transparent cover.
SPECIFICATIONS HI 8931 & HI 943500
Range
HI8931A - HI943500A
HI8931B - HI943500B
HI8931C - HI943500C
HI8931D - HI943500D
Accuracy
(@20°C/68°F)
Typical EMC
Deviation
4-20mA INPUT from
Transmitter
HI8931A
HI8931B
HI8931C
HI8931D
HI943500
Conductivity Probe
Calibration
Temp. Compensation
HI8931
HI943500
Readout
Recorder Output
1 Set Point Relay &
1 Alarm Relay
Power Supply
Panel Cutout
Environment
Weight
Enclosure