CHECKING THE RH CALIBRATION
Bring the calibration kit to a temperature
of approximately 20°C.
Remove the sintered
cap from the probe of
HI 8666.
Connect the unit to a 12 to 24VDC power
supply and hook up a multimeter (range
of 20mA) in series with the circuit.
Remove the cap from the
"RH 11.1%" chamber con-
taining the LiCl solution and
insert the HI 8666 paying at-
tention not to dip it into the
liquid.
Wait for the measurement
to stabilize (this takes ap-
proximately 4 hours).
Look at the multimeter display: the read-
ing should be 5.78 mA, representing the
theoretical chamber value of 11.1% RH.
All RH to mA conversions can be made
with the tables provided (see page 12).
Remove the HI 8666 from the chamber
and tightly seal the chamber containing
the LiCl solution.