Instruction Manual HI8564 HI 93640 Portable Thermo-Hygrometers % ins ru m en s % ins rum THER MOHY GROM ETER en s HI 85 64 therm o hygrom eter ON OFF RANG E: RH% 5 to 95 ºC 0 to ºF 32 to 60 140 RH% ºF RH% ºC ºF RAN GES HI936 ºC 40 10 ... 0 ... 95 32 ...
Dear Customer, Thank you for choosing a Hanna Instruments Product. Please read this instruction manual carefully before using the instrument. This manual will provide you with all the necessary information for the correct use of the instrument, as well as a precise idea of its versatility in a wide range of applications. These instruments are in compliance with directives.
FUNCTIONAL DESCRIPTION OF HI 8564 7 FUNCTIONAL DESCRIPTION OF HI93640 1 1 2 2 °C H 3 V e ins rum n s 8 9 10 11 °C e ins rum n s HI 8564 thermo hygrometer ON OFF 3 H 4 ON OFF 5 6 °F RH% 7 °C R ANGE S RH% 10 °C 0 °F 32 13 .... .... .... 4 5 6 RANGE: RH% 5 to 95 °C 0 to 60 °F 32 to 140 RH% 12 V THERMOHYGROMETER °F 95 60 140 °C HI93640 1) DIN socket for Relative Humidity/Temp.
SPECIFICATIONS OF HI 8564 RANGE RH °C °F 10.0 to 95.0% 0.0 to 60.0 32 to 140 RH °C °F 0.1% 0.1 1 RESOLUTION ACCURACY (@20°C / 68°F) RH °C °F +2% +0.4 +1 for one year, excluding probe error TYPICAL EMC DEVIATION RH °C °F OPERATING CONDITIONS BATTERY +2% +0.5 +1 0 to 50 oC (32 to 122 oF) 98% RH max. One 9V, 100 hours of continuous use, with low battery indicator DIMENSIONS Meter 185 x 82 x 45 mm (7.3" x 3.2" x 1.8") L:165 mm D:25 mm with 2 m cable (L:6.
OPERATIONAL GUIDE INSTRUMENT PREPARATION Each meter is supplied complete with a 9V battery. Slide off the battery compartment cover on the back of the meter (see page 12), unwrap the battery and install it. To prepare HI 8564 for use connect the RH probe to the DIN socket on the top of the meter. ON OFF Turn the meter on by pressing the ON/OFF key. TEMPERATURE MEASUREMENTS Simply press the °C or °F keys to take the temperature readings directly from the same probe.
All Hanna hygrometers have been precalibrated at the factory. Hanna Instruments uses state-of-the-art thermal humidity chambers for this purpose. It is generally recommended to have all hygrometers recalibrated at least once a year. For an accurate annual recalibration, contact your nearest Hanna Service Center. You can also check the status of your HI 8564 and perform a quick recalibration (with an accuracy of ±5%) by using the Hanna mini-calibration chambers HI 7101.
RH Calibration Procedure (HI 8564 only): • Bring the calibration kit to a temperature of approximately 20°C. • Remove the cap from the "RH 11.1%" chamber containing the LiCl solution and insert the probe paying attention not to tip it into the liquid. Li Cl 11.1% Na Cl 75.4% CALIBRATION KIT • Remove the adhesive sticker which covers the calibration trimmer access holes. • Remove the cap from the "RH 75.4%" chamber containing the NaCl solution and insert the probe.
BATTERY REPLACEMENT SHOCKPROOF RUBBER BOOTS Battery replacement must only take place in a non hazardous area using the battery types specified in this instruction manual (see page 14). FOR HI 8564 ONLY: HI 8564 When the battery is run down "V" is displayed on the LCD to warn the user. In order to replace the battery, simply slide the cover off and replace it with a new 9V battery.
ACCESSORIES HI 70601/5 HI 710001 HI 710007 HI 710008 HI 710011 HI 710031 HI 7101 HI 7111/P HI 7121/P HI 721310 MANRHR2 Detachable interchangeable TFPC RH & temp. probe for HI 8564 with 5 m (16.
OTHER PRODUCTS FROM HANNA • Chemical Test Kits • Conductivity and TDS Meters • Dissolved Oxygen Meters • Dosing Pumps • Electrodes (pH and ORP) • CE DECLARATION OF CONFORMITY DECLARATION OF CONFORMITY We Hanna Instruments Srl V.
HANNA LITERATURE HANNA LITERATURE Hanna publishes a wide range of catalogs and handbooks for an equally wide range of applications. The reference literature currently covers areas such as: • Water Treatment • Process • Swimming Pools • Agriculture • Food • Laboratory • Thermometry and many others. New reference material is constantly being added to the library. For these and other catalogs, handbooks and leaflets, contact your dealer or the Hanna Customer Service Center nearest to you.