Instruction Manual HI 8564 and HI 93640 Portable Thermo-Hygrometers % ins rum en s HI 85 64 therm o hygrom ete r RAN RH% ºC ºF GES 10 ... 0 ... 95 32 ...
Dear Customer, Thank you for choosing a HANNA product. Please read this instruction manual carefully before using the instrument. For additional technical information, do not hesitate to e-mail us at tech@hannainst.com These instruments are in compliance with the directives. WARRANTY All Hanna Instruments meters are warranted for two years against defects in workmanship and materials when used for their intended purpose and maintained according to the instructions. Probes are warranted for 6 months.
FUNCTIONAL DESCRIPTION OF HI 8564 1 2 7 °C H 12 Range Resolution 0.1% RH 0.1°C / 1°F Typical EMC Deviation 2% RH / ±0.5°C / ±1°F ins rum n s HI 8564 thermo hygrometer ON OFF 3 4 5 RH% RH Calibration Manual, 2 point, through trimmers on RH probe RH Probe (included) HI 70601/2 with built-in temperature sensor and 2 m cable Environment 0 to 50°C (32 to 122°F); RH max 98% non-condensing 6 °F °C R ANGE S 13 10.0 to 95.0% RH 0.0 to 60.0°C / 32 to 140°F Accuracy ±2% RH ±0.
FUNCTIONAL DESCRIPTION OF HI93640 1 Range 2 Resolution 3 Accuracy ±2% RH ±0.4°C / ±1°F (for 1 year excluding probe error) e °C V THERMOHYGROMETER ON OFF 4 5 6 RANGE: RH% 5 to 95 °C 0 to 60 °F 32 to 140 RH% 7 °F 5.0 to 95.0% RH 0.0 to 60.0°C / 32 to 140°C 0.1% RH 0.1°C / 1°F Typical EMC Deviation ±3% RH / ±0.4°C / ±0.8°F ins rum n s H SPECIFICATIONS OF HI 93640 Environment 0 to 50°C (32 to 122°F); RH max 98% non-condensing Battery Type / Life 1 x 9V / approx.
TEMPERATURE MEASUREMENTS Simply press the °C or °F keys to display the temperature reading. When measuring temperature in degrees Centigrade, "°C" is displayed. For any problem in taking measurements, please contact the nearest HANNA Customer Service Department. 8 The kits are composed of two thermally isolated chambers, each one equipped with a threaded cap and three bottles containing the appropriate precalibrated saturated salts to produce a known RH value. Li Cl 11.1% Na Cl 75.
l Li C • When the salt has dissolved completely, add all the content of the second bottle of HI 7111. • Bring the calibration kit to a temperature of approximately 20°C. • Remove the cap from the "RH 11.1%" chamber containing the LiCl solution and insert the probe paying attention not to tip it into the liquid. • Allow the solution obtained to cool, and then pour it into the chamber marked "RH 11.1%", making sure that no residue remains on the walls of the glass container. Li Cl 11.1% Na Cl 75.
• Remove the cap from the "RH 75.4%" chamber containing the NaCl solution and insert the probe. • Wait for the measurement to stabilize (approximately 4 hours) BATTERY REPLACEMENT Li Cl 11.1% Na Cl 75.4% CALIBRATION KIT Battery replacement must only take place in a safe area and using the battery type specified in this instruction manual. 12 9 3 6 4h • Turn the high humidity trimmer until the readout on the display is 64.3%.
ACCESSORIES HI 70601/2 HI 70601/5 HI 710001 HI 710007 HI 710008 HI 710009 HI 710010 HI 710011 HI 710031 HI 7101 HI 7102 HI 7111/P HI 7121/P CE DECLARATION OF CONFORMITY RH probe for HI 8564 with builtin temperature sensor and 2 m (6.6') cable RH probe for HI 8564 with builtin temperature sensor and 5 m (16.
SALES & TECHNICAL SERVICE Australia: Tel. (03) 9769.0666 • Fax (03) 9769.0699 China: Tel. (10) 88570068 • Fax (10) 88570060 Egypt: Tel. & Fax (02) 2758.683 Germany: Tel. (07851) 9129-0 • Fax (07851) 9129-99 Greece: Tel. (210) 823.5192 • Fax (210) 884.0210 Indonesia: Tel. (21) 4584.2941 • Fax (21) 4584.2942 Japan: Tel. (03) 3258.9565 • Fax (03) 3258.9567 Korea: Tel. (02) 2278.5147 • Fax (02) 2264.1729 Malaysia: Tel. (603) 5638.9940 • Fax (603) 5638.9829 Singapore: Tel. 6296.7118 • Fax 6291.