Owner manual

9
START UP
Place the instrument on a flat table. Do not place the instrument on direct sun light.
Connect the titrator to mains socket with ground connection and the correct voltage and
frequency. See the label on the instrument rear for this.
Place the peristaltic pump tube on the pump. See the Pump Tube Replacement section for the
procedure.
Remove the reagent bottle cap and place the bottle cap of the tubes set. Place the reagent
bottle in the appropriate place on the titrator top.
Turn the instrument ON using the power switch from the rear panel of the instrument and wait
until it displays dashes.
1) Selected mode: SETUP, CAL or CFM for confirming different values
2) Stability indicator: when reading is unstable or calibration is in progress
3) Stirrer active tags
4) Calibration messages
5) Pump active tags
6) Calibration messages
7) “pH” tag: when a buffer is displayed on the secondary display
8) Four digit and half secondary display
9) “Time” tag: when the time is displayed on the secondary display
10) Four digit and half main display
11) “pH” upper tag: when performing a pH calibration or entering SETUP mode to change the end-point
12) Automatic Temperature Compensation: when ATC blinks the temperature probe is not connected
and the temperature value will be considered to be 25
0
C.