Owner manual

15
PUMP TUBE REPLACEMENT
To remove the tube of the peristaltic pump follow next steps:
Detach the old tube system from the reagent bottle.
Grasp one fixing ring of the peristaltic pump tube.
Pull the tube until it’s taken out from its location.
Remove the other side of the tube.
To mount the new peristaltic pump tube follow next steps:
Position one peristaltic pump fixing ring on its location.
Stretch the tube over the peristaltic pump cylinders.
Fix the second pump fixing ring on its location.
Attach the tube to the reagent bottle.
Note: Purge the peristaltic pump until drops of reagent appears on the dosing tip by pressing the
PURGE button.
Dismount the tube if the titrator is not used several days.
Press START STOP in order to start the pump
calibration.
At the end of the calibration procedure done appears
for a few seconds and then the meter automatically
returns to measurement mode.