User Guide

9
GUIDE TO DISPLAY CODES
After TIMER is pressed, a blinking hourglass icon appears and the
display shows a 7 minutes coundown. Also the Zero tag might
blink if no zero measurement has been made before. At the end of
the countdown an acoustic signal alerts the user that the timer is
finished.
Indicates that the instrument is in a ready state and waiting for the
next command (Timer or Zero).
These prompts indicate the type of power supply: “Line” (if the
external power supply is used) or the battery level.
This prompt appears for a few seconds each time the instrument is
turned ON.
The instrument is zeroed and a measurement can be made.
Indicates that the meter is making a measurement.
Batteries voltage is getting low and the batteries need to be
replaced.
Indicates that the meter is performing a zero measurement. The
light intensity is automatically re-adjusted (auto-calibration features)
if necessary.
12
SAMPLE PREPARATION FOR RED WINE
To remove interference of phenols, samples of Red
Wine must be treated with activated carbon.
Fill one cuvet with 10 mL of Red Wine.
Add 2 spoons of HI 93703-59 Charcoal to the
cuvet.
Cap the cuvet and mix vigorously for 2 minutes.
Then wait for 3 minutes.
Fold a filter disc twice as shown in the figure.
Separate one side from the other three to form a
cone. Insert the folded filter disc in the funnel.
Filter the treated wine into an empty cuvet. This
is the wine sample.
Note: if the filtered wine is still red, repeat the
above procedure.
MEASUREMENT PROCEDURE