Light Detector Silicon Photocell
Environment 0 to 50°C (32 to 122°F); max 95% RH non-condensing
Battery Type 4 x 1,5 volt AA batteries / 12 to 20 VDC through voltage adapter
Auto-Shut off
After 15' of non-use.
Dimensions 225 x 85 x 80 mm (8.7 x 3.3 x 3.1")
Weight 500 g (17,6 oz.).
SPECIFICATIONS
Precision is how closely repeated measurements agree
with each other. Precision is usually expressed as
standard deviation (SD). Accuracy is defined as the
nearness of a test result to the true value.
Although good precision suggests good accuracy,
precise results can be inaccurate. The figure explains
these definitions.
For example for phenols: in a laboratory using a
standard solution of 0.350 g/L gallic acid and a
representative lot of reagent, an operator obtained
with a single instrument a standard deviation of
0.015 g/L.
PRECISION AND ACCURACY
The reaction between phenols and the Folin-Ciocalteu reagent involves oxidation of the phenolic groups (R-
OH) with a mixture of phosphotungstenic acid (H
3
PW
12
O
40
) and phosphomolybdenic acid (H
3
PMo
12
O
40
) to
the quinoid form (R=O). The concomitant reduction of the Folin-Ciocalteu reagent causes a blue color in
the sample that is proportional to the total phenolic content that, in turn, is expressed as g/L of Gallic Acid
Equivalents (GAE).
Another common method to express the phenol content is to report the Folin Index (FI) or Total Phenol
Index. Simply multiply the read value in g/L of GAE times 25, to find the Folin Index value.
24
Add 6 drops of HI 83742B-0 Folin & Ciocalteu’s
Reagent to the cuvet.
Replace the cap and shake gently to mix.
x 6
#2
(Sample)
Use the 200 µL automatic pipette with a clean
pipette tip to add exactly 0,2 mL of diluted red
wine sample.
10 mL
Wait for 1 minute and then use the 3 mL pipette to
add HI 83742C-0 Reagent C Carbonate Buffer, to
bring the
volume up to the 10 mL mark.
Turn the meter ON and press RANGE to select the
parameter code “P5” for total phenols, red wine (see
page 13). Then hold down READ/TIMER for several
seconds. The instrument will show the countdown.
Alternatively, wait for 2 hours.
The instrument gives an acoustic signal to alert the
user that the countdown is finished.
This is the reacted sample.