Instruction Manual HI 83226 Multiparameter Bench Photometer for Pool & Spa Applications www.hannainst.
Dear Customer, Thank you for choosing a Hanna product. Please read this instruction manual carefully before using the instrument. This manual will provide you with the necessary information for the correct use of the instrument. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com. TABLE OF CONTENTS PRELIMINARY EXAMINATION ..............................................................................................................................................
SIGNIFICANCE OF POOL AND SPA TESTING A major family leisure pursuit is the enjoyment of Swimming Pool and Spa facilities world-wide. A basic necessity of Pool water treatment, to ensure such enjoyment, is to maintain the water in a safe and pleasant condition for the bathers. In order to achieve such an objective, swimming pool water requires testing on daily, and sometimes hourly bases for disinfection residuals and pH.
Calcium Hardness The presence of calcium in the system is desired to ensure filming on those places where the temperature is relatively high, like in boilers and pipes transporting warm water. Scaling must be avoided because it reduces heat transfer and pump capacity. Beside the calcium carbonate deposits in the pipes, high scaling values do cause cloudy water. It is recommended to maintain the calcium hardness value within the range from 200 to 400 ppm as calcium carbonate (CaCO3).
SPECIFICATIONS Light Life Light Detector Environment Life of the instrument Silicon Photocell 0 to 50°C (32 to 122°F); max 90% RH non-condensing Power Supply external 12 Vdc power adapter Auto-Shut off built-in rechargeable battery Dimensions 235 x 200 x 110 mm (9.2 x 7.87 x 4.33") Weight 0.9 Kg For specifications related to each single method (e.g. range, resolution, etc.), refer to the related measurement section.
FUNCTIONAL DESCRIPTION KEYPAD DESCRIPTION The keypad contains 8 direct keys and 3 functional keys with the following functions: INSTRUMENT DESCRIPTION Press to perform the function displayed above it on the LCD. ESC Press to exit the current screen. Press to access the select method menu. Press to move up in a menu or a help screen, to increment a set value, to access second level functions. Press to move down in a menu or a help screen, to decrement a set value, to access second level functions.
(c) take out the syringe and clean the outside of the syringe tip. Be sure that no drops are hanging on the tip of the syringe, if so eliminate them. Then, keeping the syringe in vertical position above the cuvette, push the plunger down into the syringe until the lower edge of the seal is exactly on the 0.5 mL mark. Now the exact amount of 0.5 mL has been added to the cuvette, even if the tip still contains some solution.
specific method. • It is possible to take multiple readings in a row, but it is recommended to take a new zero reading for each sample and to use the same cuvette for zeroing and measurement when possible (for most precise results follow the measurement procedures carefully). • Discard the sample immediately after the reading is taken, or the glass might become permanently stained. • All the reaction times reported in this manual are at 25 °C (77 °F).
• After the desired method is selected, follow the measurement described in the related section. • Before performing a test you should read all the instructions carefully. DATA MANAGEMENT The instrument features a data log function to help you keep track of all your analysis. The data log can hold 200 individual measurements. Storing, viewing and deleting the data is possible using the LOG and RCL keys.. Storing data: You can store only a valid measurement.
Date / Time This option is used to set the instrument’s date and time. Press the Modify key to change the date/time. Press the ⊳ u functional keys to highlight the value to be modified (year, month, day, hour, minute or second). Use the s t keys to change the value. Press the Accept key to confirm or ESC to return to the setup without saving the new date or time. Instrument ID Time format Option: 0 to 9999. This option is used to set the instrument’s ID (identification number).
• Remove the cuvette. ALKALINITY SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0 to 500 mg/L (as CaCO3) 5 mg/L ±5 mg/L ±10% of reading at 25 °C ±5 mg/L Note: Any chlorine present in the sample will interfere with the reading. To remove the chlorine interference add one drop of HI 93755-53 Chlorine Remover to the unreacted sample. Tungsten lamp with narrow band interference filter @ 575 nm Colorimetric Method.
BROMINE • Remove the cuvette and add the content of one packet of HI 93716-0 DPD reagent. Replace the cap and shake gently for about 20 seconds to dissolve most of the reagent. SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 10.00 mg/L 0.01 mg/L ±0.08 mg/L ±3% of reading at 25 °C ±0.01 mg/L Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the Standard Methods for the Examination of Water and Wastewater, 20th edition, DPD method.
CALCIUM HARDNESS SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0 to 500 mg/L (as CaCO3) 5 mg/L ±10 mg/L ±5% of reading at 25 °C ±5 mg/L • Add 1 drop of HI 93720C-0 EGTA solution to one cuvette (# 1), replace the cap and invert the cuvette several times to mix. This is the blank. #1 #1 Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Calmagite method.
• Press the Unit key to change the current measurement unit. The results can be converted to French degrees (°f), German degrees (°dH) and English degrees (°E). FREE CHLORINE CHLORINE FREE SPECIFICATIONS Range Resolution • Press s or t to return to the measurement screen. Note: This test will detect any calcium contamination in the beaker, measuring syringes or sample cells. To test cleanliness, repeat the test multiple times until you obtain consistent results.
• Press the Zero key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement. TOTAL CHLORINE SPECIFICATIONS Range Resolution • Remove the cuvette. • Add the content of one packet of HI 93701-0 DPD reagent. Replace the cap and shake gently for 20 seconds. Accuracy Typical EMC Deviation Light Source Method • Reinsert the cuvette into the instrument. 0.00 to 5.00 mg/L 0.01 mg/L from 0.00 to 2.50 mg/L; 0.10 mg/L above 2.50 mg/L ±0.03 mg/L ±3% of reading at 25 °C ±0.
• Press the Zero key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement. FREE COPPER SPECIFICATIONS • Remove the cuvette. • Add 1 packet of HI 93711-0 DPD reagent. Replace the cap and shake gently for 20 seconds. Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 5.00 mg/L 0.01 mg/L ±0.02 mg/L ±4% of reading at 25 °C ±0.01 mg/L Tungsten lamp with narrow band interference filter @ 575 nm Adaptation of the EPA method.
• Remove the cuvette. TOTAL COPPER • Add the content of one packet of HI 93702-0 Copper Reagent. Replace the cap and shake gently for 15 seconds. • Reinsert the cuvette into the instrument. • Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait for 45 seconds and press Read. When the timer ends the meter will perform the reading. SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 5.00 mg/L 0.01 mg/L ±0.
• Remove the cuvette. CYANURIC ACID SPECIFICATIONS • Add the content of one packet of HI 93702-0 Copper Reagent. Replace the cap and shake gently for 15 seconds. • Add the content of one packet of HI 93702T-0 Copper Total Reagent. Replace the cap and shake vigorously for 15 seconds. • Replace the cuvette into the holder and ensure that the notch on the cap is positioned securely into the groove.
IRON • Reinsert the cuvette into the instrument. SPECIFICATIONS • Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait for 45 seconds and press Read. When the timer ends the meter will perform the reading. Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 5.00 mg/L 0.01 mg/L ±0.04 mg/L ±2% of reading at 25 °C ±0.
OZONE • Reinsert the cuvette into the instrument. SPECIFICATIONS • Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait for 3 minutes and press Read. When the timer ends the meter will perform the reading. Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 2.00 mg/L 0.01 mg/L ±0.02 mg/L ±3% of reading at 25 °C ±0.01 mg/L Tungsten lamp with narrow band interference filter @ 525 nm Colorimetric DPD Method.
• Press the Zero key. The display will show “-0.0-” the meter is zeroed and ready for measurement. • Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for measurement. • Remove the cuvette. • Remove the cuvette. • Add the content of one packet of HI 93757-0 Ozone Reagent. Replace the cap and shake gently for 20 seconds. • Add the content of one packet of the optional reagent HI93703-52-0 Glycine Powder. Replace the cap and shake gently until completely dissolved.
pH SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method • Remove the cuvette and add 5 drops of HI 93710-0 Phenol Red Indicator. Replace the cap and mix the solution. 6.5 to 8.5 pH 0.1 pH ±0.1 pH at 25 °C ±0.1 pH Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the Phenol Red method. The reaction with the reagent causes a yellow to red tint in the sample. • Reinsert the cuvette into the instrument.
ERRORS AND WARNINGS The instrument shows clear warning messages when erroneous conditions appear and when measured values are outside the expected range. These messages are described bellow. DATA MANAGEMENT The analyzed data can be managed using Hanna’s product HI92000, Windows® Compatible Software. No Light: The light source is not functioning properly. Light Leak: There is an excess amount of ambient light reaching the detector. Inverted cuvettes: The sample and the zero cuvettes are inverted.
ACCESSORIES REAGENT SETS HI 93701-01 100 free chlorine tests (powder) HI 93701-03 300 free chlorine tests (powder) HI 93701-F 300 free chlorine tests (liquid) HI 93703-52 Glycine Powder, Optional Reagent for 100 tests HI 93711-01 100 total chlorine tests (powder) HI 93711-03 300 total chlorine tests (powder) HI 93701-T 300 total chlorine tests (liquid) HI 93711-03 300 total chlorine tests HI 93702-01 100 free copper tests HI 93702-03 300 free copper tests HI 93702T-01 100 total copper tests HI 93702T-03 300
Hanna Instruments Inc. Highland Industrial Park 584 Park East Drive Woonsocket, RI 02895 USA Technical Support for Customers Tel. (800) 426 6287 Fax (401) 765 7575 E-mail tech@hannainst.com www.hannainst.