User Guide

28
Remove the cuvet.
Add 4 drops of the First reagent.
Replace the cap and mix the
solution by inverting the cuvet a
couple of times.
Add 4 drops of the Second
reagent. Replace the cap and
mix the solution by inverting the
cuvet a couple of times.
• Reinsert the cuvet into the
instrument.
Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 3 minutes and 30 seconds
and press READ DIRECT. In both cases “SIP” will blink during
measurement.
The instrument directly displays concentration in mg/L of ammonia
nitrogen (NH
3
-N) on the display.
To convert the reading to mg/L of ammonia (NH
3
), multiply by a
factor of 1.21.
To convert the reading to mg/L of ammonium (NH
4
+
), multiply by
a factor of 1.29.
INTERFERENCES
acetone, alcohols, aldehydes, glycine, hardness above 1 g/L, iron,
organic chloramines, sulfide, various aliphatic and aromatic amines.
x 4
Wait for a few seconds and the
display will show “-0.0-”. Now
the meter is zeroed and ready for
measurement.
x 4
Ammonia HR
33
SPECIFICATIONS
Range 0.0 to 30.0 mg/L
Resolution 0.5 mg/L
Accuracy ±1.0 mg/L ±8% of reading
Light Source Tungsten lamp with narrow band interference fil-
ter @ 525 nm
Method Adaptation of the Cadmium Reduction method.
The reaction between nitrate-nitrogen and the re-
agent causes an amber tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93728-0 Powder reagent 1 packet
REAGENT SETS
HI 93728-01 Reagents for 100 tests
HI 93728-03 Reagents for 300 tests
For other accessories see page 56.
MEASUREMENT PROCEDURE
Note: for sample preparation follow the IRRIGATION WATER (LR)
procedure at page 19.
Select the program number 4 “Nitrate LR” on the secondary
LCD by pressing PROGRAM M and L.
Using the pipette, fill the cuvet with 6mL of sample, up to
half of its height, and replace the cap.
Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO and
“SIP” will blink on
the display.
Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed
and ready for measurement.
Nitrate LR
NITRATE LOW RANGE