User Guide

26
Ammonia MR
Remove the cuvet.
Add 4 drops of the First reagent.
Replace the cap and mix the
solution by inverting the cuvet a
couple of times.
Add 4 drops of the Second
reagent. Replace the cap and mix
the solution by inverting the cuvet
a couple of times.
• Reinsert the cuvet into the
instrument.
Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 3 minutes and 30 seconds
and press READ DIRECT. In both cases “SIP” will blink during
measurement.
The instrument directly displays concentration in mg/L of ammonia
nitrogen (NH
3
-N) on the display.
To convert the reading to mg/L of ammonia (NH
3
), multiply by a
factor of 1.21.
To convert the reading to mg/L of ammonium (NH
4
+
), multiply by
a factor of 1.29.
INTERFERENCES
acetone, alcohols, aldehydes, glycine, hardness above 1 g/L, iron,
organic chloramines, sulfide, various aliphatic and aromatic amines.
x 4
Wait for a few seconds and the
display will show “-0.0-”. Now
the meter is zeroed and ready for
measurement.
x 4
35
SPECIFICATIONS
Range 0 to 150 mg/L
Resolution 2.0 mg/L
Accuracy ±5 mg/L ±8% of reading
Light Source Tungsten lamp with narrow band interference fil-
ter @ 525 nm
Method Adaptation of the Cadmium Reduction method.
The reaction between nitrate-nitrogen and the re-
agent causes an amber tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93728-0 Powder reagent 1 packet
REAGENT SETS
HI 93728-01 Reagents for 100 tests
HI 93728-03 Reagents for 300 tests
For other accessories see page 56.
MEASUREMENT PROCEDURE
Note: for sample preparation follow the NUTRIENTS SOLUTION
(MR) procedure at page 20.
Select the program number 5 “Nitrate MR” on the secondary
LCD by pressing PROGRAM M and L.
Using the pipette, add to the cuvet 6 mL of sample, up to
half of its height, and replace the cap.
Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO and “SIP”
will blink on the
display.
Wait for a few seconds and the display
will show “-0.0-”. Now the meter is
zeroed and ready for measurement.
Nitrate MR
NITRATE MEDIUM RANGE