User Guide

20
filter
disc
Mix well using the spoon
and then wait for 5
minutes.
Unscrew the filter assembly,
put a filter disc inside and
close it again.
Connect the filter assembly
to the 60 mL syringe by
the screw rim.
Take off the syringe
plunger and fill the syringe
with the sample to be
filtered. Pour the sample
gently, trying to avoid
transferring of the actived
carbon to the syringe.
Reinsert the plunger and
filter the solution into the
100mL beaker by pushing
gently the syringe plunger.
The sample is now ready.
Note: Filter at least 40 mL of solution if all the four parameters
have to be tested. If the solution is still turbid or colored, treat it
again with a packet of active carbon. After use, throw the filter
disc away and wash the syringe and the filter assembly well.
Always use a new disc for another sample.
NUTRIENTS SOLUTION (MR):
Use the graduated cylinder to measure exactly 20 mL
of sample.
41
SPECIFICATIONS
Range 0.0 to 50.0 mg/L
Resolution 0.5 mg/L
Accuracy ±2.5 mg/L ±4% of reading
Light Source Tungsten lamp with narrow band interference
filter @ 525 nm
Method Adaptation of the Standard Methods for the
Examination of Water and Wastewater, 18
th
edition, Amino Acid method. The reaction be-
tween phosphorus and reagents causes a blue
tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93706A-0 Molybdate 10 drops
HI 93706B-0 Amino Acid Powder 1 packet
REAGENT SETS
HI 93706-01 Reagents for 100 tests
HI 93706-03 Reagents for 300 tests
For other accessories see page 56.
MEASUREMENT PROCEDURE
Select the program number 8 “Phosphorus MR” on the secondary
LCD by pressing PROGRAM M and L.
Fill the cuvet with 10 mL of unreacted
sample (up to the mark) and replace
the cap.
Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO and “SIP” will blink on
the display.
Note: for sample preparation follow the NUTRIENTS SOLUTION (MR)
procedure at page 20.
Phosphorus MR
PHOSPHORUS MEDIUM RANGE
10 mL