User Guide

12
POWER CONNECTION
Plug the 12VDC adapter (HI 710005 - 110VDC, or HI 710006 -
220VDC) into the DC socket. Plug the adapter into the outlet.
Alternatively, remove the battery cover on the back of the meter;
attach 2 fresh 9V batteries and replace the cover.
Note:Insure the main line is surge protected.
Note: Always turn the meter off before unplugging it to insure no data
is lost.
MEASUREMENT PROCEDURE
Turn the meter on by pressing ON/OFF.
The meter will first perform an LCD self
diagnostic test by displaying a full set of
figures.
Then it will show a scrolling
"225 Hanna Inst" message.
OPERATIONAL GUIDE
When the LCD displays “----”, the meter
is ready. On the secondary LCD "P1" will
appear to inform that the first parameter
measurement procedure (Ammonia LR)
can be performed.
Press the PROGRAM M and PROGRAM L keys to select the
desired parameter.
49
SPECIFICATIONS
Range 20 to 200 mg/L
Resolution 5 mg/L
Accuracy ±30 mg/L ±7% of reading
Light Source Tungsten lamp with narrow band interference
filter @ 610 nm
Method Adaptation of the Turbidimetric Tetraphenylborate
method. The reaction between Potassium and
reagents causes turbidity in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93750
AA
AA
A-0 Potassium Reagent 6 drops
HI 93750
BB
BB
B-0 Powder Reagent 1 packet
REAGENT SETS
HI 93750-01 Reagents for 100 tests
HI 93750-03 Reagents for 300 tests
For other accessories see page 56.
MEASUREMENT PROCEDURE
Note: for sample preparation follow the NUTRIENTS SOLUTION (HR)
procedure at page 21.
Select the program number 12 “Potassium HR” on the secondary
LCD by pressing PROGRAM M and L.
Fill the cuvet with 10 mL of sample, up to the
mark.
Add six drops of HI 93750A,
replace the cap and swirl the
solution.
Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO adn “SIP”
will appear on the dis-
play.
10 mL
Potassium HR
POTASSIUM HIGH RANGE
x 6