Zinc
Place the cap and insert the cuvette into the instrument
and close the lid.
Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
Remove the cuvette and add 0.5 mL of HI 93731B-0
Cyclohexanone to the cuvette.
Note: To prevent any contamination from the polycarbonate
cap, prior to replacing it, close the sample cuvette with
the supplied HDPE plastic stopper.
Replace the cap and mix the sample for 15 seconds.
Insert the sample into the instrument.
Press Timer and the display will show the countdown prior to the measurement or, alternatively,
wait for 3 minutes and 30 seconds and press Read. When the timer ends the meter will perform the
reading.
The instrument displays the results in mg/L of zinc.