Manual

48
Phosphate HR
Add the content of one packet of HI 93717B-0 Phosphate
HR Reagent B to the cuvette. Replace the cap and shake
gently until completely dissolved.
Reinsert the cuvette into the instrument.
Press Timer and the display will show the countdown prior
to the measurement or, alternatively, wait for 5 minutes
and press Read. When the timer ends the meter will
perform the reading. The instrument displays the results in
mg/L of phosphate (PO
4
3
¯).
Press or to access the second level functions.
Press the Chem Frm key to convert the result in mg/L of phosphorus (P) and phosphorus pentoxide
(P
2
O
5
).
Press or to return to the measurement screen.
INTERFERENCES
Sulfide
Chloride above 150000 mg/L )
Calcium above 10000 mg/L as CaCO
3
Magnesium above 40000 mg/L as CaCO
3
Ferrous iron above 100 mg/L