Instruction Manual HI 83206 Multiparameter Bench Photometer for Environmental Testing This Instrument is in Compliance with the CE Directives www.hannainst.
Dear Customer, Thank you for choosing a Hanna product. Please read this instruction manual carefully before using the instrument. This manual will provide you with the necessary information for the correct use of the instrument. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com. This instrument is in compliance with directives. TABLE OF CONTENTS COPPER LR .................................................................... 35 CYANURIC ACID ....................
SPECIFICATIONS Life of the instrument Silicon Photocell 0 to 50°C (32 to 122°F); max 90% RH non-condensing Power Supply external 12 Vdc power adapter built-in rechargeable battery Dimensions 235 x 200 x 110 mm (9.2 x 7.87 x 4.33") Weight 0.9 Kg For specifications related to each method (e.g. range, precision, etc.) refer to the related measurement section. Light Life Light Detector Environment PRECISION AND ACCURACY Precision is how closely repeated measurements agree with each other.
FUNCTIONAL DESCRIPTION KEYPAD DESCRIPTION The keypad contains 8 direct keys and 3 functional keys with the following functions: INSTRUMENT DESCRIPTION Press to perform the function displayed above it. The functions are screen related. Press to exit the current screen. Press to access the select method menu. Press to move up in a menu or a help screen, to increment a set value, to access second level functions.
TIPS FOR AN ACCURATE MEASUREMENT The instructions listed below should be carefully followed during testing to ensure most accurate results. • Color or suspended matter in large amounts may cause interference, they should be removed by treatment with active carbon and filtration. • Ensure the cuvette is filled correctly: the liquid in the cuvette forms a convexity on the top; the bottom of this convexity must be at the same level of the 10 mL mark.
Pay attention to push the cuvette completely down in the holder. HEALTH & SAFETY The chemicals contained in the reagent kits may be hazardous if improperly handled. Read the Material Safety Data Sheet (MSDS) before performing tests. Safety equipment: Wear suitable eye protection and clothing when required, and follow instructions carefully. Reagent spills: If a reagent spill occurs, wipe up immediately and rinse with plenty of water. If reagent contacts skin, rinse the affected area thoroughly with water.
OPERATIONAL GUIDE POWER CONNECTION AND BATTERY MANAGEMENT The meter can be powered from an AC/DC adapter (included) or from the built-in rechargeable battery. Note: Always turn the meter off before unplugging it to ensure no data is lost. When the meter switches ON, it verifies if the power supply adapter is connected.
SETUP In the Setup mode the instrument’s parameters can be changed. Some parameters affect the measuring sequence and others are general parameters that change the behavior or appearance of the instrument. Press SETUP to enter the setup mode. Press ESC or SETUP to return to the main screen. A list of setup parameters will be displayed with currently configured settings. Press HELP for additional information. Press the keys to select the parameter and select a new value as follows: Backlight Values: 0 to 8.
Instrument ID Option: 0 to 9999. This option is used to set the instrument’s ID (identification number). The instrument ID is used while exchanging data with a PC. Press Modify functional key to access the instrument ID screen. Press the keys in order to set the desired value. Press Accept functional key to confirm the value or ESC to return to the setup menu without saving the new value. AMMONIA MEDIUM RANGE SPECIFICATIONS 0.00 to 10.00 mg/L 0.01 mg/L ±0.05 mg/L ±5% of reading ±0.
• Remove the cuvette. AMMONIA LOW RANGE • Add 4 drops of HI 93715A-0 First reagent (6 drops for seawater analysis). Replace the cap and mix the solution. SPECIFICATIONS x4 • Add 4 drops of HI 93715B-0 Second reagent (10 drops for seawater analysis). Replace the cap and mix the solution. Range Resolution Accuracy Typical EMC Deviation Light Source Method x4 0.00 to 3.00 mg/L 0.01 mg/L ±0.04 mg/L ±4% of reading ±0.
• Remove the cuvette. FREE CHLORINE CHLORINE FREE • Add 4 drops of HI 93700A-0 First reagent (6 drops for seawater analysis). Replace the cap and mix the solution. x4 • Add 4 drops of HI 93700B-0 Second reagent (10 drops for seawater analysis). Replace the cap and mix the solution. x4 • Reinsert the cuvette into the instrument. • Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 3 minutes and 30 seconds and press READ.
• Press ZERO key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement. • Reinsert the cuvette into the instrument. • Press READ to start the reading. The instrument displays the results in mg/L of free chlorine. • Remove the cuvette. Powder reagents procedure • Add the content of one packet of HI 93701-0 DPD reagent. Replace the cap and shake gently for 20 seconds (or 2 minutes for seawater analysis).
• Press ZERO key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement. TOTAL CHLORINE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 3.50 mg/L 0.01 mg/L ±0.03 mg/L ±3% of reading ±0.01 mg/L Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the EPA DPD method 330.5. The reaction between the chlorine and the DPD reagent causes a pink tint in the sample.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 2 minutes and 30 seconds and press READ. When the timer ends the meter will perform the reading. CHROMIUM VI HIGH RANGE SPECIFICATIONS • The instrument displays the results in mg/L of total chlorine. Note: free and total chlorine have to be measured separately with fresh unreacted samples following the related procedure if both values are requested.
CHROMIUM VI LOW RANGE • Reinsert the cuvette into the instrument. SPECIFICATIONS • Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 6 minutes and press READ. When the timer ends the meter will perform the reading. The instrument displays concentration in µg/L of chromium VI.
• Reinsert the cuvette into the instrument. • Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 6 minutes and press READ. When the timer ends the meter will perform the reading. The instrument displays concentration in µg/L of chromium VI.
• Fill the second cuvette up to the mark with unfiltered sample and replace the cap. This is the apparent color. #2 • Filter 10 mL of sample through a filter with a 0.45 µm membrane into the third cuvette, up to the 10 mL mark and replace the cap. This is the true color. • Insert the apparent color cuvette (# 2) into the instrument and close the lid. COPPER HIGH RANGE 10 mL #2 SPECIFICATIONS #3 Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 5.00 mg/L 0.01 mg/L ±0.
• Remove the cuvette. COPPER LOW RANGE • Add the content of one packet of HI 93702-0 Bicinchoninate. Replace the cap and shake gently for about 15 seconds. • Reinsert the cuvette into the instrument. • Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 45 seconds and press READ. When the timer ends the meter will perform the reading.
• Remove the cuvette. CYANURIC ACID • Add the content of one packet of HI 93747-0 Bicinchoninate. Replace the cap and shake gently for about 15 seconds. • Reinsert the cuvette into the instrument. • Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 45 seconds and press READ. When the timer ends the meter will perform the reading.
• Fill a second cuvette with 10 mL of the reacted sample up to the mark. Replace the cap. MOLYBDENUM 10 mL SPECIFICATIONS • Reinsert the cuvette into the instrument. • Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 45 seconds and press READ. When the timer ends the meter will perform the reading. Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.0 to 40.0 mg/L 0.1 mg/L ±0.3 mg/L ±5% of reading ±0.
• Fill one graduated mixing cylinder up to the 25 mL mark with the sample. • Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 5 minutes and press READ. When the timer ends the meter will perform the reading. The instrument displays concentration in mg/L of molybdenum. • Add the content of one packet of HI 93730A-0 molybdenum reagent, close the cylinder and invert it several times until completely dissolved.
• Reinsert the cuvette into the instrument. NICKEL HIGH RANGE • Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 1 minute and press READ. When the timer ends the meter will perform the reading. SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 7.00 g/L 0.01 g/L ±4% of reading ±0.02 g/L Tungsten lamp with narrow band interference filter @ 575 nm Adaptation of the photometric method.
NICKEL LOW RANGE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.000 to 1.000 mg/L 0.001 mg/L ±0.010 mg/L ±7% of reading ±0.001 mg/L Tungsten lamp with narrow band interference filter @ 575 nm Adaptation of the PAN method. The reaction between nickel and the reagents causes an orange tint in the sample.
INTERFERENCES NITRATE Interference may be caused by: Co2+ must not be present Fe2+ must not be present Al3+ above 32 mg/L Ca2+ above 1000 mg/L (as CaCO3) Cd2+ above 20 mg/L Cl– above 8000 mg/L Cr3+ above 20 mg/L Cr6+ above 40 mg/L Cu2+ above 15 mg/L F– above 20 mg/L Fe3+ above 10 mg/L K+ above 500 mg/L Mg2+ above 400 mg/L Mn2+ above 25 mg/L Mo6+ above 60 mg/L Na+ above 5000 mg/L Pb2+ above 20 mg/L Zn2+ above 30 mg/L SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.
• Replace the cap and immediately shake vigorously up and down for exactly 10 seconds. Continue to mix by inverting the cuvette gently for 50 seconds, while taking care not to induce air bubbles. Powder will not completely dissolve. Time and way of shaking could sensitively affect the measurement. • Reinsert the cuvette into the instrument, taking care not to shake it.
NITRITE LOW RANGE • Reinsert the cuvette into the instrument. SPECIFICATIONS • Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 10 minutes and press READ. When the timer ends the meter will perform the reading. The instrument displays concentration in mg/L of nitrite. Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 0.35 mg/L 0.01 mg/L ±0.02 mg/L ±4% of reading ±0.
DISSOLVED OXYGEN • Reinsert the cuvette into the instrument. SPECIFICATIONS • Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 6 minutes and press READ. When the timer ends the meter will perform the reading. The instrument displays concentration in mg/L of nitrite. Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.0 to 10.0 mg/L 0.1 mg/L ±0.4 mg/L ±3% of reading ± 0.
pH • Let the sample stand and the flocculent agent will start to settle. • After approximately 2 minutes, when the upper half of the bottle becomes limpid, add 10 drops of HI 93732C-0. x 10 • Replace the cap and invert the bottle until the settled flocculent dissolves completely. The sample is ready for measurement when it is yellow and completely limpid. SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 6.5 to 8.5 pH 0.1 pH ±0.1 pH ±0.
PHOSPHATE HIGH RANGE • Reinsert the cuvette into the instrument. SPECIFICATIONS • Press the READ key to start the reading. The instrument displays the pH value. Range Resolution Accuracy Typical EMC Dev. Light Source Method 0.0 to 30.0 mg/L 0.1 mg/L ±1 mg/L ±4% of reading ±0.1 mg/L Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Amino Acid method.
• Add the content of one packet of HI 93717B-0 Phosphate HR Reagent B to the cuvette. Replace the cap and shake gently until completely dissolved. • Reinsert the cuvette into the instrument. • Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 5 minutes and press READ. When the timer ends the meter will perform the reading. The instrument displays the results in mg/L of phosphate (PO43-).
PHOSPHORUS • Reinsert the cuvette into the instrument. SPECIFICATIONS • Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 3 minutes and press READ. When the timer ends the meter will perform the reading. The instrument displays concentration in mg/L of phosphate (PO43-). Range Resolution Accuracy Typical EMC Dev. Light Source Method 0.0 to 15.0 mg/L 0.1 mg/L ±0.3 mg/L ±4% of reading ±0.
• Add the content of one packet of HI 93706B-0 Phosphorus Reagent B (Amino Acid) to the cuvette. Replace the cap and shake gently until completely dissolved. • Reinsert the cuvette into the instrument. • Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 5 minutes and press READ. When the timer ends the meter will perform the reading. The instrument displays the results in mg/L of phosphorus (P).
INTERFERENCES • Place the cuvette into the holder and close the lid. • Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement. Interference may be caused by: Phosphate above 60 mg/L Phosphate above 75 mg/L Sulfide and high concentration of iron Eliminate color and turbidity interferences by zeroing the meter with the original water sample.
SILVER SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method • Add exactly 1.0 mL of HI 93737B-0 Buffer reagent to the second beaker (the sample) and swirl gently to mix. 0.000 to 1.000 mg/L 0.001 mg/L ±0.005 mg/L ±10% of reading ± 0.001 mg/L #2 Tungsten lamp with narrow band interference filter @ 575 nm. Adaptation of the PAN method. The reaction between silver and reagents causes an orange tint in the sample.
• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement. ZINC SPECIFICATIONS 10 mL • Fill a second cuvette up to the mark with 10 mL of the reacted sample. #2 • Insert the second cuvette into the instrument. Range Resolution Accuracy Typical EMC Deviation Light Source Method #2 • Press READ to start the reading. The instrument displays the results in mg/L of silver. 0.00 to 3.00 mg/L 0.01 mg/L ±0.03 mg/L ±3% of reading ±0.
• Place the cap and insert the cuvette into the instrument and close the lid. ERRORS AND WARNINGS • Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement. The instrument shows clear warning messages when erroneous conditions appear and when measured values are outside the expected range. These messages are described below. No Light: The light source is not functioning properly. • Remove the cuvette and add 0.5 mL of HI 93731B-0 cyclohexanone to the cuvette.
DATA MANAGEMENT The analyzed data can be managed using Hanna’s product HI92000, Windows® Compatible Software. Phosphorus Silica Silver Zinc 0.0 to 15.0 mg / L 0.00 to 2.00 mg/L 0.000 to 1.000 mg/L 0.00 to 3.00 mg/L ACCESSORIES STANDARD METHODS Description Ammonia MR Ammonia LR Chlorine, Free Chlorine, Total Chromium VI HR Chromium VI LR Color of Water Copper HR Copper LR Cyanuric Acid Molybdenum Nickel HR Nickel LR Nitrate Nitrite HR Nitrite LR Oxygen, Dissolved pH Phosphate HR Phosphate LR Range 0.
HI 93732-01 HI 93732-03 HI 93737-01 HI 93737-03 HI 93740-01 HI 93740-03 HI 93747-01 HI 93747-03 HI 93749-01 HI 93749-03 100 dissolved oxygen tests 300 dissolved oxygen tests 50 silver tests 150 silver tests 50 nickel LR tests 150 nickel LR tests 100 copper LR tests 300 copper LR tests 100 chromium VI LR tests 300 chromium VI LR tests WARRANTY OTHER ACCESSORIES HI 731318 cloth for wiping cuvettes (4 pcs) HI 731321 glass cuvettes (4 pcs) HI 731325W new cap for cuvette (4 pcs) HI 740034 cap for 100 mL beake
SALES AND TECHNICAL SERVICE CONTACTS Australia: Tel. (03) 9769.0666 • Fax (03) 9769.0699 e-mail: hannains@hannainst.com.au China: Tel. (10) 88570068 • Fax (10) 88570060 e-mail: hannachina@vip.sina.com Egypt: Tel. & Fax (02) 2758.683 e-mail: hannaegypt@go.com.eg Germany: Tel. (07851) 9129-0 • Fax (07851) 9129-99 e-mail: hannager@aol.com Greece: Tel. (210) 823.5192 • Fax (210) 884.0210 e-mail: hannagr@otenet.gr Indonesia: Tel. (21) 4584.2941 • Fax (21) 4584.2942 e-mail: transit@dnet.net.