Instruction Manual HI 83205 Multiparameter Bench Photometer for Boilers and Cooling Towers This Instrument is in Compliance with the CE Directives www.hannainst.
Dear Customer, Thank you for choosing a Hanna product. Please read this instruction manual carefully before using the instrument. This manual will provide you with the necessary information for the correct use of the instrument. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com. This instrument is in compliance with directives. TABLE OF CONTENTS PRELIMINARY EXAMINATION ............................................. 3 ABBREVIATIONS ..............................
SPECIFICATIONS Life of the instrument Silicon Photocell 0 to 50°C (32 to 122°F); max 90% RH non-condensing Power Supply external 12 Vdc power adapter built-in rechargeable battery Dimensions 235 x 200 x 110 mm (9.2 x 7.87 x 4.33") Weight 0.9 Kg For specifications related to each method (e.g. range, precision, etc.) refer to the related measurement section. Light Life Light Detector Environment PRECISION AND ACCURACY Precision is how closely repeated measurements agree with each other.
FUNCTIONAL DESCRIPTION KEYPAD DESCRIPTION The keypad contains 8 direct keys and 3 functional keys with the following functions: INSTRUMENT DESCRIPTION Press to perform the function displayed above it. The functions are screen related. ESC Press to exit the current screen. Press to access the select method menu. Press to move up in a menu or a help screen, to increment a set value, to access second level functions.
TIPS FOR AN ACCURATE MEASUREMENT The instructions listed below should be carefully followed during testing to ensure most accurate results. • Color or suspended matter in large amounts may cause interference, they should be removed by treatment with active carbon and filtration. • Ensure the cuvette is filled correctly: the liquid in the cuvette forms a convexity on the top; the bottom of this convexity must be at the same level of the 10 mL mark.
Pay attention to push the cuvette completely down in the holder. HEALTH & SAFETY The chemicals contained in the reagent kits may be hazardous if improperly handled. Read the Material Safety Data Sheet (MSDS) before performing tests. Safety equipment: Wear suitable eye protection and clothing when required, and follow instructions carefully. Reagent spills: If a reagent spill occurs, wipe up immediately and rinse with plenty of water. If reagent contacts skin, rinse the affected area thoroughly with water.
OPERATIONAL GUIDE • After the desired method was selected, follow the measurement described in the related section. • Before performing a test you should read all the instructions carefully. POWER CONNECTION AND BATTERY MANAGEMENT The meter can be powered from an AC/DC adapter (included) or from the built-in rechargeable battery. Note: Always turn the meter off before unplugging it to ensure no data is lost. When the meter switches ON, it verifies if the power supply adapter is connected.
SETUP In the Setup mode the instrument’s parameters can be changed. Some parameters affect the measuring sequence and others are general parameters that change the behavior or appearance of the instrument. Press SETUP to enter the setup mode. Press ESC or SETUP to return to the main screen. A list of setup parameters will be displayed with currently configured settings. Press HELP for additional information. Press the keys to select the parameter and select a new value as follows: Backlight Values: 0 to 8.
Instrument ID Option: 0 to 9999. This option is used to set the instrument’s ID (identification number). The instrument ID is used while exchanging data with a PC. Press Modify functional key to access the instrument ID screen. Press the keys in order to set the desired value. Press Accept functional key to confirm the value or ESC to return to the setup menu without saving the new value.
• Add the content of one packet of HI 93712C-0 Bleaching Powder to one of the two cuvettes. Replace the cap and shake vigorously until completely dissolved. This is the blank. AMMONIA MEDIUM RANGE #1 SPECIFICATIONS • Place the blank into the holder and close the lid. • Press TIMER and the display will show the countdown prior to zeroing the blank. Alternatively wait for 15 minutes and then press ZERO. The display will show "-0.0-" when the meter is zeroed and ready for measurement.
• Remove the cuvette. AMMONIA LOW RANGE • Add 4 drops of HI 93715A-0 First reagent (6 drops for seawater analysis). Replace the cap and mix the solution. SPECIFICATIONS x4 • Add 4 drops of HI 93715B-0 Second reagent (10 drops for seawater analysis). Replace the cap and mix the solution. Range Resolution Accuracy Typical EMC Deviation Light Source Method x4 0.00 to 3.00 mg/L 0.01 mg/L ±0.04 mg/L ±4% of reading ±0.
• Remove the cuvette. BROMINE • Add 4 drops of HI 93700A-0 First reagent (6 drops for seawater analysis). Replace the cap and mix the solution. SPECIFICATIONS x4 • Add 4 drops of HI 93700B-0 Second reagent (10 drops for seawater analysis). Replace the cap and mix the solution. x4 • Reinsert the cuvette into the instrument. • Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 3 minutes and 30 seconds and press READ.
• Remove the cuvette and add the content of one packet of HI 93716-0 DPD reagent. Replace the cap and shake gently for about 20 seconds to dissolve most of the reagent. FREE CHLORINE CHLORINE FREE SPECIFICATIONS • Reinsert the cuvette into the instrument. • Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 2 minutes and 30 seconds and press READ. When the timer ends the meter will perform the reading.
• Press ZERO key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement. • Reinsert the cuvette into the instrument. • Press READ to start the reading. The instrument displays the results in mg/L of free chlorine. • Remove the cuvette. Powder reagents procedure • Add the content of one packet of HI 93701-0 DPD reagent. Replace the cap and shake gently for 20 seconds (or 2 minutes for seawater analysis).
• Press ZERO key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement. TOTAL CHLORINE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 3.50 mg/L 0.01 mg/L ±0.03 mg/L ±3% of reading ±0.01 mg/L Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the EPA DPD method 330.5. The reaction between the chlorine and the DPD reagent causes a pink tint in the sample.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 2 minutes and 30 seconds and press READ. When the timer ends the meter will perform the reading. CHLORINE DIOXIDE SPECIFICATIONS • The instrument displays the results in mg/L of total chlorine. Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 2.00 mg/L 0.01 mg/L ±0.10 mg/L ±5% of reading ±0.
• Add the content of one packet of HI 93738B-0 Dechlorinating Reagent to one of the two cylinders (#1), close and invert it several times until it is totally dissolved. This is the blank. • Press ZERO key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement. #1 • Add precisely 0.5 mL of HI 93738C-0 Chlorine Dioxide Reagent to each cylinder (#1 & #2), close them and invert several times to mix. #2 10 mL #2 • Insert the sample into the instrument.
CHROMIUM VI HIGH RANGE • Reinsert the cuvette into the instrument. SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0 to 1000 µg/L 1 µg/L ±5 µg/L ±4% of reading ±1 µg/L • Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 6 minutes and press READ. When the timer ends the meter will perform the reading. The instrument displays concentration in µg/L of chromium VI.
CHROMIUM VI LOW RANGE • Reinsert the cuvette into the instrument. • Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 6 minutes and press READ. When the timer ends the meter will perform the reading. The instrument displays concentration in µg/L of chromium VI.
• Remove the cuvette. COPPER HIGH RANGE • Add the content of one packet of HI 93702-0 Bicinchoninate. Replace the cap and shake gently for about 15 seconds. SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 5.00 mg/L 0.01 mg/L ±0.02 mg/L ±4% of reading ±0.01 mg/L Tungsten lamp with narrow band interference filter @ 575 nm Adaptation of the EPA method. The reaction between copper and the bicinchoninate reagent causes a purple tint in the sample.
• Remove the cuvette. COPPER LOW RANGE • Add the content of one packet of HI 93747-0 Bicinchoninate. Replace the cap and shake gently for about 15 seconds. SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0 to 1000 μg/L 1 μg/L ±10 μg/L ±5% of reading ±1 μg/L Tungsten lamp with narrow band interference filter @ 575 nm Adaptation of the EPA method. The reaction between copper and the bicinchoninate reagent causes a purple tint in the sample.
HYDRAZINE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0 to 400 µg/L 1 µg/L ±4% of full scale ±2 µg/L Tungsten lamp with narrow band interference filter @ 420 nm Adaptation of the ASTM Manual of Water and Environmental Technology, method D1385-88, p-Dimethylaminobenzaldehyde method. The reaction between hydrazine and the liquid reagent causes a yellow tint in the sample. REQUIRED REAGENT Code HI 93704-0 • Add 12 drops of the HI 93704-0 reagent to each cuvette.
IRON HIGH RANGE • Reinsert the cuvette into the instrument. SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 5.00 mg/L 0.01 mg/L ±0.04 mg/L ±2% of reading ±0.01 mg/L Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the EPA Phenantroline method 315B, for natural and treated waters. The reaction between iron and reagents causes an orange tint in the sample.
• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement. IRON LOW RANGE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0 to 400 µg/L 1 µg/L ±10 µg/L ±8% of reading ±1 µg/L • Remove the cuvette. Tungsten lamp with narrow band interference filter @ 575 nm Adaptation of the TPTZ Method. The reaction between iron and the reagent causes a violet tint in the sample.
INTERFERENCES Interference may be caused by: Cadmium above 4.0 mg/L Chromium3+ above 0.25 mg/L Chromium6+ above 1.2 mg/L Cobalt above 0.05 mg/L Copper above 0.6 mg/L Cyanide above 2.8 mg/L Manganese above 50.0 mg/L Mercury above 0.4 mg/L Molybdenum above 4.0 mg/L Nickel above 1.0 mg/L Nitrite ion above 0.8 mg/L Sample pH should be between 3 and 4 to avoid developed color to fade or turbidity formation. MOLYBDENUM SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.
• Fill one graduated mixing cylinder up to the 25 mL mark with the sample. • Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for five minutes and press READ. When the timer ends the meter will perform the reading. The instrument displays concentration in mg/L of molybdenum. • Add the content of one packet of HI 93730A-0 molybdenum reagent, close the cylinder and invert it several times until completely dissolved.
NITRATE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.0 to 30.0 mg/L 0.1 mg/L ±0.5 mg/L ±10% of reading ±0.1 mg/L Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the cadmium reduction method. The reaction between nitrate and the reagent causes an amber tint in the sample. REQUIRED REAGENTS Code HI 93728-0 Description Powder reagent Quantity 1 packet • Replace the cap and immediately shake vigorously up and down for exactly 10 seconds.
NITRITE HIGH RANGE • Reinsert the cuvette into the instrument. SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0 to 150 mg/L 1 mg/L ±4 mg/L ±4% of reading ±1 mg/L Tungsten lamp with narrow band interference filter @ 575 nm Adaptation of the Ferrous Sulfate method. The reaction between nitrite and the reagent causes a greenish-brown tint in the sample.
NITRITE LOW RANGE • Reinsert the cuvette into the instrument. SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 0.35 mg/L 0.01 mg/L ±0.02 mg/L ±4% of reading ±0.01 mg/L Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the EPA Diazotization method 354.1. The reaction between nitrite and the reagent causes a pink tint in the sample.
DISSOLVED OXYGEN • Let the sample stand and the flocculent agent will start to settle. SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method • After approximately 2 minutes, when the upper half of the bottle becomes limpid, add 10 drops of HI 93732C-0. 0.0 to 10.0 mg/L 0.1 mg/L ±0.4 mg/L ±3% of reading ± 0.
pH • Reinsert the cuvette into the instrument. SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 6.5 to 8.5 pH 0.1 pH ±0.1 pH ±0.1 pH • Press the READ key to start the reading. The instrument displays the pH value. Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the Phenol Red method. The reaction with the reagent causes a yellow to red tint in the sample.
• Add the content of one packet of HI 93717B-0 Phosphate HR Reagent B to the cuvette. Replace the cap and shake gently until completely dissolved. PHOSPHATE HIGH RANGE SPECIFICATIONS Range Resolution Accuracy Typical EMC Dev. Light Source Method 0.0 to 30.0 mg/L 0.1 mg/L ±1 mg/L ±4% of reading ±0.1 mg/L Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Amino Acid method.
PHOSPHATE LOW RANGE • Reinsert the cuvette into the instrument. SPECIFICATIONS Range Resolution Accuracy Typical EMC Dev. Light Source Method 0.00 to 2.50 mg/L 0.01 mg/L ±0.04 mg/L ±4% of reading ±0.01 mg/L Tungsten lamp with narrow band interference filter @ 610 nm Adaptation of the Ascorbic Acid method. The reaction between phosphate and the reagent causes a blue tint in the sample.
SILICA • Place the cuvette into the holder and close the lid. SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method • Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement. 0.00 to 2.00 mg/L 0.01 mg/L ±0.03 mg/L ±3% of reading ±0.01 mg/L Tungsten lamp with narrow band interference filter @ 610 nm Adaptation of the ASTM Manual of Water and Environmental Technology, D859, Heteropoly Blue method.
INTERFERENCES Interference may be caused by: Phosphate above 60 mg/L Phosphate above 75 mg/L Sulfide and high concentration of iron Eliminate color and turbidity interferences by zeroing the meter with the original water sample. ZINC SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 3.00 mg/L 0.01 mg/L ±0.03 mg/L ±3% of reading ±0.
• Place the cap and insert the cuvette into the instrument and close the lid. ERRORS AND WARNINGS • Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement. The instrument shows clear warning messages when erroneous conditions appear and when measured values are outside the expected range. These messages are described below. No Light: The light source is not functioning properly. • Remove the cuvette and add 0.5 mL of HI 93731B-0 cyclohexanone to the cuvette.
DATA MANAGEMENT The analyzed data can be managed using Hanna’s product HI92000, Windows® Compatible Software. pH Phosphate HR Phosphate LR Silica Zinc 6.5 to 8.5 pH 0.0 to 30.0 mg/L 0.00 to 2.50 mg/L 0.00 to 2.00 mg/L 0.00 to 3.
HI 93730-01 HI 93730-03 HI 93731-01 HI 93731-03 HI 93732-01 HI 93732-03 HI 93746-01 HI 93746-03 HI 93747-01 HI 93747-03 HI 93749-01 HI 93749-03 100 molybdenum tests 300 molybdenum tests 100 zinc tests 300 zinc tests 100 dissolved oxygen tests 300 dissolved oxygen tests 50 iron LR tests 150 iron LR tests 100 copper LR tests 300 copper LR tests 100 chromium VI LR tests 300 chromium VI LR tests WARRANTY OTHER ACCESSORIES HI 731318 cloth for wiping cuvettes (4 pcs) HI 731321 glass cuvettes (4 pcs) HI 731325W
SALES AND TECHNICAL SERVICE CONTACTS Australia: Tel. (03) 9769.0666 • Fax (03) 9769.0699 e-mail: hannains@hannainst.com.au China: Tel. (10) 88570068 • Fax (10) 88570060 e-mail: hannachina@vip.sina.com Egypt: Tel. & Fax (02) 2758.683 e-mail: hannaegypt@go.com.eg Germany: Tel. (07851) 9129-0 • Fax (07851) 9129-99 e-mail: hannager@aol.com Greece: Tel. (210) 823.5192 • Fax (210) 884.0210 e-mail: hannagr@otenet.gr Indonesia: Tel. (21) 4584.2941 • Fax (21) 4584.2942 e-mail: transit@dnet.net.