Potassium LR
POTASSIUM LOW RANGE
SPECIFICATIONS
Range 0.0 to 20.0 mg/L
Resolution 0.5 mg/L
Accuracy ±3.0 mg/L ±7% of reading at 25 °C
Typical EMC ±0.5 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 610 nm
Method Adaptation of the Turbidimetric Tetraphenylborate method. The reaction between
Potassium and reagents causes turbidity in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93750A-0 Potassium Reagent 6 drops
HI 93750B-0 Powder Reagent 1 packet
REAGENT SETS
HI 93750-01 Reagents for 100 tests
HI 93750-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
Note: for sample preparation follow the COLORED OR TURBID SAMPLES procedure at page 17.
Select the
Potassium LR
method using the procedure
described in the
Method Selection
section (see page 12).
Fill the cuvette with 10 mL of sample, up to the mark.
Add 6 drops of HI 93750A-0 Potassium Reagent,
replace the cap and swirl the solution.
Place the cuvette into the holder and close the lid.
Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
Remove the cuvette and add the content of one packet of
HI 93750B-0 reagent. Replace the cap and gently mix for
one minute by slowly turning the cuvette upside down.
Reinsert the cuvette into the instrument.
Press Timer and the display will show the countdown prior to the
measurement or, alternatively, wait for 2 minutes and press Read.
When the timer ends the meter will perform the reading. The
instrument displays the results in mg/L (ppm) of potassium (K).
Press s or t to access the second level functions.
Press the Chem Frm key to convert the result in mg/L of potassium oxide (K
2
O).
Press s or t to return to the measurement screen.
INTERFERENCES
Interferences may be caused by:
Ammonium above 10 ppm
Calcium above 10000 ppm as CaCO
3
Chloride above 12000 ppm
Magnesium above 8000 ppm as CaCO
3
Sodium above 8000 ppm
Potassium MR
10 mL