User Manual

100 101
pH
Remove the cuvette and add 5 drops of HI 93710-0 Phenol
Red Indicator. Replace the cap and mix the solution.
Reinsert the cuvette into the instrument.
Press. the Read key to start the reading. The instrument displays the pH value.
pH
pH
SPECIFICATIONS
Range 6.5 to 8.5 pH
Resolution 0.1 pH
Accuracy ±0.1 pH at 25 °C
Typical EMC ±0.1 pH
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 525 nm
Method Adaptation of the Phenol Red method. The reaction with the reagent causes a yellow
to red tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93710-0 Phenol Red Indicator 5 drops
REAGENT SETS
HI 93710-01 Reagents for 100 pH tests
HI 93710-03 Reagents for 300 pH tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
Select the
pH
method using the procedure described in the
Method Selection
section (see page 12).
Fill the cuvette with 10 mL of unreacted sample (up to the
mark) and replace the cap.
Place the cuvette into the holder and close the lid.
Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
10 mL