User Manual

84 85
Nickel HR
Remove the cuvette and add the content of one packet of
HI 93726-0 reagent. Replace the cap and shake gently
until completely dissolved.
Reinsert the cuvette into the instrument.
Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait
for 1 minute and press Read. When the timer ends the meter will perform the reading.
The instrument displays concentration in g/L of nickel.
INTERFERENCES
Interference may be caused by copper.
Nickel HR
SPECIFICATIONS
Range 0.00 to 7.00 g/L
Resolution 0.01 g/L
Accuracy ±0.07 ±4% of reading at 25 °C
Typical EMC ±0.02 g/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 575 nm
Method Adaptation of the photometric method. The reaction between nickel and the reagent
causes a blue tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93726-0 Powder reagent 1 packet
REAGENT SETS
HI 93726-01 Reagents for 100 tests
HI 93726-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
Select the
Nickel HR
method using the procedure described in the
Method Selection
section (see page 12).
Fill the cuvette up to the mark with 10 mL of unreacted sample
and replace the cap.
Place the cuvette into the holder and close the lid.
Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
NICKEL HIGH RANGE
10 mL