80 81 Molybdenum
MOLYBDENUM
SPECIFICATIONS
Range 0.0 to 40.0 mg/L
Resolution 0.1 mg/L
Accuracy ±0.3 mg/L ±5% of reading at 25 °C
Typical EMC ±0.1 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 420 nm
Method Adaptation of the mercaptoacetic acid method. The reaction between molybdenum and
the reagents causes a yellow tint in the sample.
REQUIRED REAGENT
Code Description Quantity
HI 93730A-0 Reagent A 1 packet
HI 93730B-0 Reagent B 1 packet
HI 93730C-0 Reagent C 1 packet
REAGENT SETS
HI 93730-01 Reagents for 100 tests
HI 93730-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
Select the
Molybdenum
method using the procedure described
in the
Method Selection
section (see page 12).
Fill the cuvette with 10 mL of unreacted sample (up to the
mark) and replace the cap.
Place the cuvette into the holder and close the lid.
Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
Manganese LR
Insert the second cuvette with the reacted sample into the instrument.
Press Read to start the reading. The instrument displays the results in
µµ
µµ
µg/L of manganese.
Press s or t to access the second level functions.
Press the Chem Frm key to convert the result in µg/L of potassium permanganate (KMnO
4
) and
permanganate (MnO
4
).
Press s or t to return to the measurement screen.
INTERFERENCES
Interference may be caused by:
Aluminum above 20 mg/L
Cadmium above 10 mg/L
Calcium above 200 mg/L as CaCO
3
Cobalt above 20 mg/L
Copper above 50 mg/L
Iron above 10 mg/L
Lead above 0.5 mg/L
Magnesium above 100 mg/L as CaCO
3
Nickel above 40 mg/L
Zinc above 15 mg/L.
10 mL