User Manual

78 79
Manganese LR
Add 0.2 mL of the HI 93748B-0 Alkaline-cyanide reagent
solution to each cuvette, replace the caps and invert gently
to mix for about 30 seconds.
Add 1 mL of the HI 93748C-0 0.1% PAN indicator
solution to each cuvette, replace the caps and shake
gently.
Add 3 drops of Dispersing Agent (HI 93703-51) to each
cuvette, replace the caps and invert gently to mix for
about 30 seconds.
Place the cuvette with the reacted deionized water (blank)
into the holder and close the lid.
Press Timer and the display will show the countdown prior to zeroing the blank. Alternatively wait for
2 minutes and then press Zero. The display will show "-0.0-" when the meter is zeroed and ready for
measurement.
Manganese LR
MANGANESE LOW RANGE
SPECIFICATIONS
Range 0 to 300 µg/L
Resolution 1 µg/L
Accuracy ±10 µg/L ±3% of reading at 25 °C
Typical EMC ±1 µg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 575 nm
Method Adaptation of the PAN Method. The reaction between manganese and the reagents
causes an orange tint in the sample.
REQUIRED REAGENT
Code Description Quantity
HI 93748A-0 Ascorbic acid 2 packets
HI 93748B-0 Alkaline-cyanide sol. 0.40 mL
HI 93748C-0 0.1% PAN indicator 2 mL
HI 93703-51 Dispersing Agent 4-6 drops
REAGENT SETS
HI 93748-01 Reagents for 50 tests
HI 93748-03 Reagents for 150 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
Select the
Manganese LR
method using the procedure described in the
Method Selection
section (see
page 12).
Fill one cuvette with 10 mL of deionized water (up to the mark).
Fill a second cuvette with 10 mL of sample (up to the mark).
Add the content of one packet of HI 93748A-0 Ascorbic acid to
each cuvette, replace the caps and shake gently until completely
dissolved.
10 mL
10 mL