User Manual

38 39
Chlorine Dioxide
CHLORINE DIOXIDE
SPECIFICATIONS
Range 0.00 to 2.00 mg/L
Resolution 0.01 mg/L
Accuracy ±0.10 mg/L ±5% of reading at 25 °C
Typical EMC ±0.01 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 575 nm
Method Adaptation of the Chlorophenol Red method. The reaction between chlorine dioxide and
reagents causes a colorless to purple tint in the sample.
REQUIRED REAGENT
Code Description Quantity
HI 93738A-0 Reagent A 1 mL
HI 93738B-0 Dechlorinating Reagent B 1 packet
HI 93738C-0 Reagent C 1 mL
HI 93738D-0 Reagent D 1 mL
REAGENT SETS
HI 93738-01 Reagents for 100 tests
HI 93738-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
Select the
Chlorine Dioxide
method using the procedure
described in the
Method Selection
section (see page 12).
Fill two graduated mixing cylinders (#1 & #2) up to the
25 mL mark with the sample.
Add 0.5 mL of HI 93738A-0 Chlorine Dioxide Reagent to each cylinder (#1 & #2), close them and
invert several times to mix.
Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait
for 2 minutes and 30 seconds and press Read. When the timer ends the meter will perform the
reading.
The instrument displays the results in mg/L of total chlorine.
Note: free and total chlorine have to be measured separately with fresh unreacted samples following
the related procedure if both values are requested.
INTERFERENCES
Interference may be caused by: Bromine, Iodine, Ozone, Oxidized forms of Chromium and Manganese.
In case of water with hardness greater than 500 mg/L CaCO
3
, shake the sample for approximately
2 minutes after adding the powder reagent.
In case of water with alkalinity greater than 250 mg/L CaCO
3
or acidity greater than 150 mg/L CaCO
3
,
the color of the sample may develop only partially, or may rapidly fade. To resolve this, neutralize the
sample with diluted HCl or NaOH.
Total Chlorine
# 2
25 mL
# 1
25 mL
# 1 & # 2
# 1 # 2