SPECIFICATIONS
Range 0 to 500 mg/L (as CaCO
3
)
Resolution 5 mg/L
Accuracy ±5 mg/L ±10 % of reading at 25 °C
Typical EMC ±5 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 575 nm
Method Colorimetric Method. The reaction causes a distinctive range of colors from yellow to
green to greenish blue to develop.
REQUIRED REAGENTS
Code Description Quantity/test
HI 93755-0 Alkalinity Indicator Reagent 1 mL
REAGENT SETS
HI 93755-01 Reagents for 100 tests
HI 93755-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
Select the
Alkalinity
method using the procedure described in the
Method Selection
section (see page 12).
Fill the cuvette with 10 mL of unreacted sample (up to the mark)
and replace the cap.
Place the cuvette into the holder and close the lid
Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
ALKALINITY
Alkalinity
Add the content of one packet of HI 93712C-0 Bleaching
powder to one of the two cuvettes. Replace the cap and shake
vigorously until completely dissolved. This is the blank.
Place the blank into the holder and close the lid.
Press Timer and the display will show the countdown prior to
zeroing the blank. Alternatively wait for 15 minutes and then press
Zero. The display will show "-0.0-" when the meter is zeroed and
ready for measurement.
Remove the blank and insert the other cuvette into the instrument.
Press the Read key and the meter will perform the reading. The instrument displays the results in
mg/L of aluminum.
Press s or t to access the second level functions.
Press the Chem Frm key to convert the result in mg/L of Al
2
O
3
.
Press s or t to return to the measurement screen.
INTERFERENCES
Interference may be caused by:
Iron above 20 mg/L, Alkalinity above 1000 mg/L, Phosphate above 50 mg/L; Fluoride must be absent.
Aluminum
# 1
# 1
# 2
10 mL