Instruction Manual HI 83200 Multiparameter Bench Photometer for Laboratories www.hannainst.
Dear Customer, Thank you for choosing a Hanna product. Please read this instruction manual carefully before using the instrument. This manual will provide you with the necessary information for the correct use of the instrument. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com. TABLE OF CONTENTS PRELIMINARY EXAMINATION ............................................. 3 GENERAL DESCRIPTION ..................................................... 3 ABBREVIATIONS ....
ABBREVIATIONS EPA: °C: °F: µg/L: mg/L: g/L: mL: HR: MR: LR: PAN: TPTZ: US Environmental Protection Agency degree Celsius degree Fahrenheit micrograms per liter (ppb) milligrams per liter (ppm) grams per liter (ppt) milliliter high range medium range low range 1-(2-pyridylazo)-2-naphtol 2,4,6-tri-(2-pyridyl)-1,3,5-triazine SPECIFICATIONS Light Life Light Detector Environment Life of the instrument Silicon Photocell 0 to 50°C (32 to 122°F); max 90% RH non-condensing Power Supply external 12 Vdc power adapt
The measurement process is carried out in two phases: first the meter is zeroed and then the actual measurement is performed. The cuvette has a very important role because it is an optical element and thus requires particular attention. It is important that both the measurement and the calibration (zeroing) cuvette are optically identical to provide the same measurement conditions. Most methods use the same cuvette for both, so it is important that measurements are taken at the same optical point.
TIPS FOR AN ACCURATE MEASUREMENT The instructions listed below should be carefully followed during testing to ensure most accurate results. • Color or suspended matter in large amounts may cause interference, and should be removed by treatment with active carbon and filtration: refere to Sample Preparation Chapter (page 17). • Ensure the cuvette is filled correctly: the liquid in the cuvette forms a convexity on the top; the bottom of this convexity must be at the same level as the 10 mL mark.
• Pay attention to push the cuvette completely down in the holder and to align the white point on the cap to the indexing mark on the meter. HEALTH & SAFETY • The chemicals contained in the reagent kits may be hazardous if improperly handled. • Read the Material Safety Data Sheet (MSDS) before performing tests. • Safety equipment: Wear suitable eye protection and clothing when required, and follow instructions carefully.
OPERATIONAL GUIDE POWER CONNECTION AND BATTERY MANAGEMENT The meter can be powered from an AC/DC adapter (included) or from the built-in rechargeable battery. Note: Always turn the meter off before unplugging it to ensure no data is lost. When the meter switches ON, it verifies if the power supply adapter is connected.
SETUP In the Setup mode the instrument’s parameters can be changed. Some parameters affect the measuring sequence and others are general parameters that change the behavior or appearance of the instrument. Press SETUP to enter the setup mode. Press ESC or SETUP to return to the main screen. A list of setup parameters will be displayed with currently configured settings. Press HELP for additional information. Press the s t keys to select a parameter and change the value as follows: Backlight Values: 0 to 8.
Instrument ID Option: 0 to 9999. This option is used to set the instrument’s ID (identification number). The instrument ID is used while exchanging data with a PC. Press the Modify key to access the instrument ID screen. Press the s t keys in order to set the desired value. Press the Accept key to confirm the value or ESC to return to the setup menu without saving the new value.
• Connect the filter assembly to the 60 mL syringe by the screw rim. • Remove the cap and fill the Demineralizer Bottle with tap water. • Remove the syringe plunger and fill the syringe with the sample to be filtered. Pour the sample gently, trying to avoid transferring of the actived carbon to the syringe. • Replace the cap and shake gently for at least 2 minutes. • Reinsert the plunger and filter the solution into the 100 mL beaker by pushing gently the syringe plunger. The sample is now ready.
• If the solution contains some turbidity or color, add a powder packet of active carbon and follow the procedure described in previous section Colored or Turbid Samples. • Calculate the dilution factor N: N = 100/V Where: V is the volume of original sample poured in the cylinder, expressed in mL, and 100 is the final volume in the cylinder, expressed in mL.
• Add the content of one packet of HI 93712C-0 Bleaching powder to one of the two cuvettes. Replace the cap and shake vigorously until completely dissolved. This is the blank. ALKALINITY SPECIFICATIONS #1 • Place the blank into the holder and close the lid. • Press Timer and the display will show the countdown prior to zeroing the blank. Alternatively wait for 15 minutes and then press Zero. The display will show "-0.0-" when the meter is zeroed and ready for measurement.
• Remove the cuvette. AMMONIA MEDIUM RANGE SPECIFICATIONS Note: Any chlorine present in the sample will interfere with the reading. To remove the chlorine interference add one drop of HI 93755-53 Chlorine Remover to the unreacted sample • Carefully add exactly 1 mL of HI 93755-0 Liquid Alkalinity Reagent using the supplied syringe. Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 10.00 mg/L 0.01 mg/L ±0.05 mg/L ±5% of reading at 25 °C ±0.
• Remove the cuvette. AMMONIA LOW RANGE SPECIFICATIONS • Add 4 drops of HI 93715A-0 First Reagent (6 drops for seawater analysis). Replace the cap and mix the solution. • Add 4 drops of HI 93715B-0 Second Reagent (10 drops for seawater analysis). Replace the cap and mix the solution. Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 3.00 mg/L 0.01 mg/L ±0.04 mg/L ±4% of reading at 25 °C ±0.
• Remove the cuvette. BROMINE SPECIFICATIONS • Add 4 drops of HI 93700A-0 First Reagent (6 drops for seawater analysis). Replace the cap and mix the solution. • Add 4 drops of HI 93700B-0 Second Reagent (10 drops for seawater analysis). Replace the cap and mix the solution. • Reinsert the cuvette into the instrument. Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 8.00 mg/L 0.01 mg/L ±0.08 mg/L ±3% of reading at 25 °C ±0.
• Remove the cuvette and add the content of one packet of HI 93716-0 DPD reagent. Replace the cap and shake gently for about 20 seconds to dissolve most of the reagent. CALCIUM SPECIFICATIONS Range Resolution Accuracy Light Source Method • Reinsert the cuvette into the instrument. 0 to 400 mg/L 10 mg/L ±10 mg/L ±5% of reading at 25 °C Tungsten lamp with narrow band interference filter @ 466 nm Adaptation of the Oxalate method.
• Replace the cap and invert several times to mix. FREE CHLORINE CHLORINE FREE • Place the cuvette into the holder and close the lid. • Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for measurement. • Remove the cuvette. • Using the 1 mL syringe, add exactly 1 mL of the HI 93752B-0 Calcium Oxalate Reagent. Replace the cap and invert the cuvette 10 times to mix (about 15 seconds). • Press Timer or wait for 5 minutes.
• Press the Zero key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement. • Reinsert the cuvette into the instrument. • Remove the cuvette. • Press Read to start the reading. The instrument displays the results in mg/L of free chlorine. Powder reagents procedure • Add the content of one packet of HI 93701 DPD reagent. Replace the cap and shake gently for 20 seconds (or 2 minutes for seawater analysis). • Reinsert the cuvette into the instrument.
• Press the Zero key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement. TOTAL CHLORINE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 3.50 mg/L 0.01 mg/L ±0.03 mg/L ±3% of reading at 25 °C ±0.01 mg/L Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the EPA DPD method 330.5. The reaction between the chlorine and the DPD reagent causes a pink tint in the sample.
• Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait for 2 minutes and 30 seconds and press Read. When the timer ends the meter will perform the reading. • The instrument displays the results in mg/L of total chlorine. CHLORINE DIOXIDE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 2.00 mg/L 0.01 mg/L ±0.10 mg/L ±5% of reading at 25 °C ±0.
• Add the content of one packet of HI 93738B-0 Dechlorinating Reagent to one of the two cylinders (#1), close and invert it several times until it is totally dissolved. This is the blank. • Press the Zero key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement. #1 #2 • Add precisely 0.5 mL of HI 93738C-0 Chlorine Dioxide Reagent to each cylinder (#1 & #2), close them and invert several times to mix.
CHROMIUM VI HIGH RANGE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0 to 1000 µg/L 1 µg/L ±5 µg/L ±4% of reading at 25 °C ±1 µg/L • Reinsert the cuvette into the instrument. Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the ASTM Manual of Water and Environmental Technology, D1687-92, Diphenylcarbohydrazide method. The reaction between chromium VI and the reagent causes a purple tint in the sample.
CHROMIUM VI LOW RANGE • Reinsert the cuvette into the instrument. SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0 to 300 µg/L 1 µg/L ±1 µg/L ±4% of reading at 25 °C ±1 µg/L Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the ASTM Manual of Water and Environmental Technology, D1687-92, Diphenylcarbohydrazide method. The reaction between chromium VI and the reagent causes a purple tint in the sample.
• Fill the second cuvette up to the mark with unfiltered sample and replace the cap. This is the apparent color. COLOR OF WATER 10 mL SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method #2 0 to 500 PCU (Platinum Cobalt Units) 1 PCU ±10 PCU ±5% of reading at 25 °C ± 1 PCU • Filter 10 mL of sample through a filter with a 0.45 µm membrane into the third cuvette, up to the 10 mL mark and replace the cap. This is the true color.
• Remove the cuvette. COPPER HIGH RANGE • Add the content of one packet of HI 93702-0 Bicinchoninate. Replace the cap and shake gently for about 15 seconds. SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 5.00 mg/L 0.01 mg/L ±0.02 mg/L ±4% of reading at 25 °C ±0.01 mg/L • Reinsert the cuvette into the instrument. Tungsten lamp with narrow band interference filter @ 575 nm Adaptation of the EPA method.
• Remove the cuvette. COPPER LOW RANGE • Add the content of one packet of HI 93747-0 Bicinchoninate. Replace the cap and shake gently for about 15 seconds. SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0 to 1000 µg/L 1 µg/L ±10 µg/L ±5% of reading at 25 °C ±1 µg/L • Reinsert the cuvette into the instrument. Tungsten lamp with narrow band interference filter @ 575 nm Adaptation of the EPA method.
Note: Pay attention to the way the spoon is filled: - do not press the powder; - do not overfill it. CYANIDE SPECIFICATIONS Range Resolution Accuracy Typical EMC Dev. Light Source Method 0.000 to 0.200 mg/L 0.001 mg/L ±0.005 mg/L ±3% of reading at 25 °C ±0.001 mg/L Tungsten lamp with narrow band interference filter @ 610 nm Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Pyridine-Pyrazolone method.
• Press s or t to access the second level functions. • Press the Chem Frm key to convert the result in mg/L of Potassium Cyanide (KCN). CYANURIC ACID SPECIFICATIONS • Press s or t to return to the measurement screen. Note: for most accurate results perform the test at 20-25 °C. Range Resolution Accuracy Typical EMC Deviation Light Source Method INTERFERENCES REQUIRED REAGENTS Interference may be caused by large amounts of turbidity that will cause high readings.
FLUORIDE • Add the content of one packet of HI 93722-0 Cyanuric Acid Reagent. Replace the cap and shake gently for about 10 seconds (dissolution is not complete). • Reinsert the cuvette into the instrument. • Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait for 45 seconds and press Read. When the timer ends the meter will perform the reading. SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 2.00 mg/L 0.
• Press Timer and the display will show the countdown prior to zeroing the blank or, alternatively, wait for two minutes and press Zero. The display will show “-0.0-” when the meter is zeroed and ready for measurement. CALCIUM HARDNESS SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 2.70 mg/L 0.01 mg/L ±0.11 mg/L ±5% of reading at 25 °C ±0.
• Press the Unit key to change the current measurement unit. The results can be converted to French degrees (°f), German degrees (°dH) and English degrees (°E). • Add 1 drop of HI 93720C-0 EGTA solution to one cuvette (# 1), replace the cap and invert the cuvette several times to mix. This is the blank. #1 #1 • Press s or t to return to the measurement screen. • Place the blank (# 1) into the holder and close the lid. • Press the Zero key. The meter will show “-0.
MAGNESIUM HARDNESS • Fill both cuvettes up to the 10 mL mark. 10 mL 10 mL #1 #2 SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 2.00 mg/L 0.01 mg/L ±0.11 mg/L ±5% of reading at 25 °C ±0.02 mg/L Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, EDTA colorimetric method. The reaction between magnesium and reagents causes a reddish-violet tint in the sample.
• Press s or t to access the second level functions. • Press the Chem Frm key to convert the result in mg/L of Magnesium (Mg). HYDRAZINE SPECIFICATIONS • Press the Unit key to change the current measurement unit. The results can be converted to French degrees (°f), German degrees (°dH) and English degrees (°E).
IODINE SPECIFICATIONS • Add 12 drops of the HI 93704-0 reagent to each cuvette. Replace the caps and shake gently to mix (about 30 seconds). #1 #2 #1 • Place the blank (#1) into the holder and close the lid. Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.0 to 12.5 mg/L 0.1 mg/L ±0.1 mg/L ±5% of reading at 25 °C ±0.
IRON HIGH RANGE • Reinsert the cuvette into the instrument. SPECIFICATIONS • Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait for 2 minutes and 30 seconds and press Read. When the timer ends the meter will perform the reading. Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 5.00 mg/L 0.01 mg/L ±0.04 mg/L ±2% of reading at 25 °C ±0.
IRON LOW RANGE • Reinsert the cuvette into the instrument. SPECIFICATIONS • Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait for 3 minutes and press Read. When the timer ends the meter will perform the reading. Range Resolution Accuracy Typical EMC Deviation Light Source Method 0 to 400 µg/L 1 µg/L ±10 µg/L ±8% of reading at 25 °C ±1 µg/L Tungsten lamp with narrow band interference filter @ 575 nm Adaptation of the TPTZ Method.
• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for measurement. • Remove the cuvette. 25 mL • Fill another graduated mixing cylinder up to the 25 mL mark with the sample. 12 9 • Add the content of one packet of HI 93746-0 TPTZ reagent, close the cylinder and shake vigorously for 30 seconds. This is the reacted sample. 30" 3 6 INTERFERENCES Interference may be caused by: Cadmium above 4.0 mg/L Chromium3+ above 0.25 mg/L Chromium6+ above 1.2 mg/L Cobalt above 0.
MAGNESIUM • Remove the cuvette. SPECIFICATIONS Range Resolution Accuracy Light Source Method 0 to 150 mg/L 5 mg/L ±5 mg/L ±3% of reading at 25 °C Tungsten lamp with narrow band interference filter @ 466 nm Adaptation of the Calmagite method. 0.5 mL of sample • Using the other 1 mL syringe, add to the cuvette exactly 0.5 mL of sample. Note: Do not mix up the two syringes! • Replace the cap and invert several times to mix. • Reinsert the cuvette into the instrument.
MANGANESE HIGH RANGE • Add the content of one packet of HI 93709B-0 Sodium Periodate reagent. Replace the cap and invert to mix with gently movements for 2 minutes. SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.0 to 20.0 mg/L 0.1 mg/L ±0.2 mg/L ±3% of reading at 25 °C ±0.1 mg/L • Reinsert the cuvette into the instrument.
MANGANESE LOW RANGE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0 to 300 µg/L 1 µg/L ±10 µg/L ±3% of reading at 25 °C ±1 µg/L • Add 0.2 mL of the HI 93748B-0 Alkaline-cyanide reagent solution to each cuvette, replace the caps and invert gently to mix for about 30 seconds. Tungsten lamp with narrow band interference filter @ 575 nm Adaptation of the PAN Method. The reaction between manganese and the reagents causes an orange tint in the sample.
MOLYBDENUM • Insert the second cuvette with the reacted sample into the instrument. SPECIFICATIONS • Press Read to start the reading. The instrument displays the results in µg/L of manganese. Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.0 to 40.0 mg/L 0.1 mg/L ±0.3 mg/L ±5% of reading at 25 °C ±0.1 mg/L Tungsten lamp with narrow band interference filter @ 420 nm Adaptation of the mercaptoacetic acid method.
• Fill one graduated mixing cylinder up to the 25 mL mark with the sample. 25 mL • Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait for five minutes and press Read. When the timer ends the meter will perform the reading. The instrument displays concentration in mg/L of molybdenum. • Add the content of one packet of HI 93730A-0 molybdenum reagent, close the cylinder and invert it several times until completely dissolved.
NICKEL HIGH RANGE • Remove the cuvette and add the content of one packet of HI 93726-0 reagent. Replace the cap and shake gently until completely dissolved. SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 7.00 g/L 0.01 g/L ±0.07 ±4% of reading at 25 °C ±0.02 g/L • Reinsert the cuvette into the instrument. Tungsten lamp with narrow band interference filter @ 575 nm Adaptation of the photometric method.
NICKEL LOW RANGE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.000 to 1.000 mg/L 0.001 mg/L ±0.010 mg/L ±7% of reading at 25 °C ±0.001 mg/L Tungsten lamp with narrow band interference filter @ 575 nm Adaptation of the PAN method. The reaction between nickel and the reagents causes an orange tint in the sample. Description Phthalate-phosphate 0.3% PAN indicator EDTA Dispersing Agent • Fill one cuvette up to the mark with 10 mL of the blank.
INTERFERENCES NITRATE Interference may be caused by: Co2+ must not be present Fe2+ must not be present Al3+ above 32 mg/L Ca2+ above 1000 mg/L (as CaCO3) Cd2+ above 20 mg/L Cl¯ above 8000 mg/L Cr3+ above 20 mg/L Cr6+ above 40 mg/L Cu2+ above 15 mg/L F¯ above 20 mg/L Fe3+ above 10 mg/L K+ above 500 mg/L Mg2+ above 400 mg/L Mn2+ above 25 mg/L Mo6+ above 60 mg/L Na+ above 5000 mg/L Pb2+ above 20 mg/L Zn2+ above 30 mg/L SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.
• Replace the cap and immediately shake vigorously up and down for exactly 10 seconds. Continue to mix by inverting the cuvette gently for 50 seconds, while taking care not to induce air bubbles. Powder will not completely dissolve. Time and way of shaking could sensitively affect the measurement. • Reinsert the cuvette into the instrument, taking care not to shake it.
NITRITE LOW RANGE • Add the content of one packet of HI 93708-0 reagent. Replace the cap and shake gently until completely dissolved. • Reinsert the cuvette into the instrument. • Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait for 10 minutes and press Read. When the timer ends the meter will perform the reading. The instrument displays concentration in mg/L of nitrite.
DISSOLVED OXYGEN • Reinsert the cuvette into the instrument. SPECIFICATIONS • Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait for 6 minutes and press Read. When the timer ends the meter will perform the reading. The instrument displays concentration in mg/L of nitrite. Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.0 to 10.0 mg/L 0.1 mg/L ±0.4 mg/L ±3% of reading at 25 °C ± 0.
OZONE • Let the sample stand and the flocculent agent will start to settle. SPECIFICATIONS • After approximately 2 minutes, when the upper half of the bottle becomes limpid, add 10 drops of HI 93732C-0. Range Resolution Accuracy Typical EMC Deviation Light Source Method • Replace the cap and invert the bottle until the settled flocculent dissolves completely. The sample is ready for measurement when it is yellow and completely limpid.
• Press the Zero key. The display will show “-0.0-” the meter is zeroed and ready for measurement. • Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for measurement. • Remove the cuvette. • Remove the cuvette. • Add the content of one packet of HI 93757-0 Ozone Reagent. Replace the cap and shake gently for 20 seconds. • Add the content of one packet of the optional reagent HI 93703-52-0 Glycine Powder. Replace the cap and shake gently until completely dissolved.
pH SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 6.5 to 8.5 pH 0.1 pH ±0.1 pH at 25 °C ±0.1 pH • Remove the cuvette and add 5 drops of HI 93710-0 Phenol Red Indicator. Replace the cap and mix the solution. Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the Phenol Red method. The reaction with the reagent causes a yellow to red tint in the sample.
PHOSPHATE HIGH RANGE SPECIFICATIONS Range Resolution Accuracy Typical EMC Dev. Light Source Method 0.0 to 30.0 mg/L 0.1 mg/L ±1 mg/L ±4% of reading at 25 °C ±0.1 mg/L Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Amino Acid method. The reaction between phosphate and reagents causes a blue tint in the sample. Description Molybdate Reagent B • Reinsert the cuvette into the instrument.
PHOSPHATE LOW RANGE SPECIFICATIONS Range Resolution Accuracy Typical EMC Dev. Light Source Method 0.00 to 2.50 mg/L 0.01 mg/L ±0.04 mg/L ±4% of reading at 25 °C ±0.01 mg/L Tungsten lamp with narrow band interference filter @ 610 nm Adaptation of the Ascorbic Acid method. The reaction between phosphate and the reagent causes a blue tint in the sample. • Reinsert the cuvette into the instrument.
PHOSPHORUS SPECIFICATIONS Range Resolution Accuracy Typical EMC Dev. Light Source Method 0.0 to 15.0 mg/L 0.1 mg/L ±0.3 mg/L ±4% of reading at 25 °C ±0.2 mg/L Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Amino Acid method. The reaction between phosphate and reagents causes a blue tint in the sample. Description Molybdate Amino Acid Powder • Reinsert the cuvette into the instrument.
• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for measurement. POTASSIUM HIGH RANGE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 20 to 200 mg/L 5 mg/L ±30 mg/L ±7% of reading at 25 °C ±5 mg/L Tungsten lamp with narrow band interference filter @ 610 nm Adaptation of the Turbidimetric Tetraphenylborate method. The reaction between Potassium and reagents causes turbidity in the sample.
INTERFERENCES POTASSIUM MEDIUM RANGE Interferences may be caused by: Ammonium above 10 ppm Calcium above 10000 ppm as CaCO3 Chloride above 12000 ppm Magnesium above 8000 ppm as CaCO3 Sodium above 8000 ppm SPECIFICATIONS POTASSIUM ULTRA HIGH RANGE For samples containing more than 200 ppm of Potassium: follow the sample preparation procedure described at page 18 for CONCENTRATED SAMPLES.
• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for measurement. POTASSIUM LOW RANGE SPECIFICATIONS • Remove the cuvette and add the content of one packet of HI 93750B-0 reagent. Replace the cap and gently mix for one minute by slowly turning the cuvette upside down. • Reinsert the cuvette into the instrument. 1’ 12 9 3 6 Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.0 to 20.0 mg/L 0.5 mg/L ±3.0 mg/L ±7% of reading at 25 °C ±0.
• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for measurement. SILICA SPECIFICATIONS • Remove the cuvette and add the content of one packet of HI 93750B-0 reagent. Replace the cap and gently mix for one minute by slowly turning the cuvette upside down. 1’ 12 9 3 6 Range Resolution Accuracy Typical EMC Deviation Light Source Method • Reinsert the cuvette into the instrument. 0.00 to 2.00 mg/L 0.01 mg/L ±0.03 mg/L ±3% of reading at 25 °C ±0.
• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for measurement. SILVER SPECIFICATIONS • Remove the cuvette and add the content of one packet of HI 93705C-0 Amino acid reagent and shake until it is completely dissolved. • Reinsert the cuvette into the instrument. • Press Timer and the display will show the countdown prior to the measurement. Alternatively, wait for exactly 3 minutes and press Read. When the timer ends the meter will perform the reading.
• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for measurement. • Add exactly 1.0 mL of HI 93737B-0 Buffer reagent to the second beaker (the sample) and swirl gently to mix. Press Timer and the display will show the countdown prior to adding reagent C or, alternatively, wait for 2 minutes. #2 • Fill a second cuvette up to the mark with 10 mL of the reacted sample. 10 mL #2 • Then add exactly 1.0 mL of HI 93737C-0 Indicator reagent to each beaker and swirl.
SULFATE SPECIFICATIONS Range Resolution Accuracy Light Source Method 0 to 100 mg/L 5 mg/L ±5 mg/L ±3% of reading at 25 °C Tungsten lamp with narrow band interference filter @ 466 nm Sulfate is precipitated with barium chloride crystals. Light absorbance of the suspension is measured. • Add the content of one packet of HI 93751-0 Indicator reagent. 1’ 12 9 3 6 • Replace the cap and invert gently to mix for 1 minute (about 30 inversions). • Reinsert the cuvette into the instrument.
ZINC • Place the cap and insert the cuvette into the instrument and close the lid. SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 3.00 mg/L 0.01 mg/L ±0.03 mg/L ±3% of reading at 25 °C ±0.01 mg/L • Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
INTERFERENCES ERRORS AND WARNINGS Interference may be caused by: Aluminum above 6 mg/L Cadmium above 0.5 mg/L Copper above 5 mg/L Iron above 7 mg/L Manganese above 5 mg/L Nickel above 5 mg/L The instrument shows clear warning messages when erroneous conditions appear and when measured values are outside the expected range. These messages are described below. No Light: The light source is not functioning properly. Light Leak: There is an excess amount of ambient light reaching the detector.
DATA MANAGEMENT The analyzed data can be managed using Hanna’s product HI92000, Windows ® Compatible Software. Windows ® is registered Trademark of "Microsoft Co.
ACCESSORIES REAGENT SETS HI 93700-01 100 ammonia LR tests HI 93700-03 300 ammonia LR tests HI 93701-01 100 free chlorine tests (powder) HI 93701-03 300 free chlorine tests (powder) HI 93701-F 300 free chlorine tests (liquid) HI 93701-T 300 total chlorine tests (liquid) HI 93702-01 100 copper HR tests HI 93702-03 300 copper HR tests HI 93704-01 100 hydrazine tests HI 93704-03 300 hydrazine tests HI 93705-01 100 silica tests HI 93705-03 300 silica tests HI 93706-01 100 phosphorus tests HI 93706-03 300 phospho
OTHER ACCESSORIES HI 731318 cloth for wiping cuvettes (4 pcs) HI 731321 glass cuvettes (4 pcs) HI 731325W new cap for cuvette (4 pcs) HI 740034 cap for 100 mL beaker (6 pcs) HI 740036 100 mL plastic beaker (6 pcs) HI 740038 60 mL glass bottle and stopper HI 740142 1 mL graduated syringe HI 740143 1 mL graduated syringe (6 pcs) HI 740144 pipette tip (6 pcs) HI 740157 plastic refilling pipette (20 pcs) HI 740220 25 mL glass cylinders with caps (2 pcs) HI 740223 170 mL plastic beaker HI 740224 170 mL plastic b
Hanna Instruments Inc. Highland Industrial Park 584 Park East Drive Woonsocket, RI 02895 USA Technical Support for Customers Tel. (800) 426 6287 Fax (401) 765 7575 E-mail tech@hannainst.com www.hannainst.