User Guide

7
Temperature Calibration
The HI 8064 thermohygrometer is factory cali-
brated and therefore usually it requires no fur-
ther intervention by the user.
If recalibration is desired, proceed as follows:
Remove the adhesive sticker which covers
the calibration trimmer access holes.
Insert the probe into a recipient in which air
heated at a temperature of 50 ºC has been
forced to circulate.
Insert a thermometer with an accuracy of 0.1
ºC into the same recipient.
Read the temperature value on the thermom-
eter.
Adjust the temperature calibration trimmer
(the central trimmer) until the measured tem-
perature value is displayed on the LCD.
TAKING MEASUREMENTS
For a rapid response, expose the end of the
humidity detector to an air current moving at
more than 0.5 m/second.
In absence of air movement, the response
can be speeded up by shaking the probe.
If condensed drops form upon the surface of
the humidity sensors, turn the instrument off
and wait until this undesired water has com-
pletely evaporated. To speed up the evapora-
tion process, expose the humidity sensors to
an air current.
If the meter is used in dusty surroundings, the
filter must be kept on at all time.