User Guide

6
Calibration procedure
Bring the calibration kit to a temperature of
approximately 20 ºC.
Remove the plug from the chamber contain-
ing the LiCl solution and insert the probe.
Remove the adhesive sticker which covers
the calibration trimmer access holes.
Wait for the reading to stabilize (this will take
a few hours) and then turn the "Low humidity
trimmer" (the one nearest to the probe con-
nection cable) until the value of 11.1% RH will
be displayed on the LCD.
Remove the probe and tightly seal the cham-
ber containing the LiCl solution.
Remove the plug from the second chamber
and insert the probe.
Wait for the reading to stabilize and then note
the displayed value.
By using the Calibration Table (pages 8-9),
find the displayed readout value on column
"A" and the corresponding RH value of col-
umn "B", to which the meter has to be brought
by acting on the "High humidity trimmer".
Adjust the "High humidity trimmer" (the one
opposed to the connection cable) until the
required value is displayed on the LCD.
Finally adjust the "Low humidity trimmer" until
the value of 75.4% RH will be displayed on the
LCD.