Instruction Manual HI 8064 Portable Thermohygrometer This Instrument is in Compliance with the CE Directives 1
Dear Customer, Thank you for choosing a Hanna product. Please read this instruction manual carefully before using the instrument. This manual will provide you with all the necessary information for the correct use of the instrument, as well as a more precise idea of its versatility in a wide range of applications. This instrument is in compliance with the directives. TABLE OF CONTENTS Preliminary Examination ............................ 3 General Description ....................................
PRELIMINARY EXAMINATION Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipment. If there is any damage, immediately notify your Dealer. Note: Save all packing material until you are sure that the instrument functions correctly. All defective items must be returned in their original packing with the supplied accessories.
SPECIFICATIONS Range Resolution Accuracy Response time Calibration Display Environment Battery Type Battery Life Dimensions Shipping weight RH 10.0 to 95.0 % Temp. 0.0 to 60.0 ºC RH 0.1 % Temp. 0.1 ºC RH ±2% Temp. ±0.4 ºC 6 seconds for 95% of accuracy through 3 trimmers in the probe • • 4-digit LCD + symbols Battery wear indicator 0 to 50°C (32 to 122°F) Max.
CALIBRATION Relative Humidity Calibration The HI 8064 thermohygrometer is factory calibrated, but it is recommended to periodically control the calibration status of the meter. Use the HI 7601 calibration kit composed of a container with two separate chambers, each one equipped with a threaded plug and three envelops containing the necessary salts. Preparing the calibration solutions • Pour approximately 26 cc of distilled water into a glass container.
Calibration procedure • Bring the calibration kit to a temperature of approximately 20 ºC. • Remove the plug from the chamber containing the LiCl solution and insert the probe. • Remove the adhesive sticker which covers the calibration trimmer access holes. • Wait for the reading to stabilize (this will take a few hours) and then turn the "Low humidity trimmer" (the one nearest to the probe connection cable) until the value of 11.1% RH will be displayed on the LCD.
Temperature Calibration The HI 8064 thermohygrometer is factory calibrated and therefore usually it requires no further intervention by the user. If recalibration is desired, proceed as follows: • Remove the adhesive sticker which covers the calibration trimmer access holes. • Insert the probe into a recipient in which air heated at a temperature of 50 ºC has been forced to circulate. • Insert a thermometer with an accuracy of 0.1 ºC into the same recipient. • Read the temperature value on the thermometer.
RELATIVE HUMIDITY CALIBRATION TABLE 8
WARRANTY All Hanna Instruments meters are warranted for two years against defects in workmanship and materials when used for their intended purpose and maintained according to the instructions. The probes are warranted for a period of six months. Damages due to accident, misuse, tampering or lack of prescribed maintenance are not covered. This warranty is limited to repair or replacement free of charge. If service is required, contact the dealer from whom you purchased the instrument.
HANNA PRODUCTS • CALIBRATION AND MAINTENANCE SOLUTIONS • CHEMICAL TEST KITS • CHLORINE METERS • CONDUCTIVITY/TDS METERS • DISSOLVED OXYGEN METERS • HYGROMETERS • ION SPECIFIC METERS (Colorimeters) • MAGNETIC STIRRERS • Na/NaCl METERS • pH/ORP/Na ELECTRODES • PROBES (DO, µS/cm, RH, T, TDS) • PUMPS • REAGENTS • SOFTWARE • THERMOMETERS • TITRATORS • TRANSMITTERS • TURBIDITY METERS • Wide Range of Accessories Most Hanna meters are available in the following formats: • BENCH-TOP METERS • POCKET-SIZE METERS • POR
MAN8064 07/01 Stay in Touch with the Manufacturer! 12