User Guide
16
• Adjust the calibration trimmer (see page 7,#6) on the side of the
instrument with the calibration screwdriver until the display shows
the TDS reading at 25°C (see the TDS vs. temperature chart on
page 19). E.g. @25°C "1382 mg/L".
• All subsequent measurements will be compensated to 25°C (77°F).
If you prefer to standardize the temperature compensation to
20°C (68°F) rather than 25°C (77°F), leave the TEMPERATURE
knob to 18°C (if the temperature of the solution is 18°C). Adjust
the trimmer to read "1251 mg/L" (see the TDS vs. temperature
chart on page 19). All subsequent measurements will be compen-
sated to 20°C.
• The calibration is now complete and the instrument is ready for
use.
The instrument should be re-calibrated once a month, or when the
probe or battery is changed.
Note: For more accurate results, it is advisable to use a calibration
solution close to the range to be measured. See the accessories
section on page 23 for a wide selection of TDS solutions.
PROCEDURE FOR HI 933000
• Pour sufficient quantity of a conductivity
calibration solution (e.g. HI 7030/
HI 8030) into a beaker to cover the holes
on the probe. If possible, use plastic bea-
kers to minimize any EMC interference.
• Immerse the conductivity probe in the
solution, making sure that holes are com-
pletely submerged.
• Wait for a couple of minutes for thermal
equilibrium to be reached.
• Tap the probe on the bottom, then shake it while rotating to make
sure no air bubbles remain trapped in the sleeve.
T
E
M
P
E
R
A
T
U
R
E
HI 7030