User Guide

14
PROCEDURE FOR HI 8733
Pour sufficient quantity of a conductivity
calibration solution (e.g. HI 7030/
HI 8030) into a beaker to cover the holes
on the probe. If possible, use plastic bea-
kers to minimize any EMC interference.
Immerse the conductivity probe in the
solution, making sure that holes are com-
pletely submerged.
Wait for a couple of minutes for thermal
equilibrium to be reached.
Tap the probe on the bottom, then shake it while rotating to
make sure no air bubbles remain trapped in the sleeve.
Switch the instrument on by pressing
ON/OFF.
Set the temperature coefficient knob to
2% to compensate for the temperature
effect of average solutions (to determine
exact value for a particular solution, see
page 20).
Select 19.99 mS/cm range by pressing
the appropriate range key.
Adjust the calibration trimmer (see page 6, #5) on the side of
the instrument with the calibration screwdriver until the display
shows "12.88 mS" i.e. the conductivity reading @ 25°C.
All subsequent measurements will be compensated to 25°C (77°F).
If you prefer to standardize the temperature compensation to
20°C (68°F) rather than 25°C (77°F), adjust the trimmer to read
"11.67 mS" (see the conductivity vs. temperature chart on page
18). All subsequent measurements will be compensated to 20°C.
The calibration is now complete and the instrument is ready for
use.
HI 7030
ON
OFF
2.5
2
1.51
0.5
0
TEMPERATURE COEFFICIENT
19.99
mS
m
S
mS