With HI 8633 and HI 8734 press the COND/TEMP key (for
HI 8633) or TDS/TEMP (for HI 8734) to display the temperature
and adjust the temperature knob to that of the solution e.g. 20°C.
Select the appropriate measurement range.
Note: If the display shows only a "1" on the
far left hand side, the meter is out of
range. Select the next (higher) range.
For HI 8733:
Adjust the TEMPERATURE COEFFICIENT
knob to 2% to compensate for the tem-
perature effect of average solutions (to
determine exact value for a particular so-
lution, see page 20).
For HI 8733 and HI 933000:
Wait for a couple of minutes for the temperature sensor to reach
thermal equilibrium with the sample before taking measure-
ments.
When the sample's temperature is lower than 20°C or higher
than 30°C, allow more time for the thermal equilibrium of the
system to be achieved.
After the measurement has been completed, the instrument should
be switched off and the probe should be cleaned and dried (see
"Probe Maintenance" on page 21).