Owner manual

76
The instrument should be re-calibrated at least once a month, or
after probe or battery replacement.
For more accurate results, it is recommended to use a calibration
solution with a conductivity value close to the range to be measured.
See the "Accessories" section for a wide selection of conductivity
solutions.
PROCEDURE FOR HI 8033
Pour the conductivity calibration solution
(e.g. HI 7030) into a beaker while mak-
ing sure that the holes on the probe are
completely submerged. If possible, use
plastic beakers to minimize any EMC in-
terference.
Immerse the conductivity probe and the
ChecktempC in the solution.
Wait a couple of minutes for the thermal
equilibrium to be reached.
Tap the probe on the bottom, then shake it while rotating to
make sure no air bubbles remain trapped in the sleeve.
Measure the temperature of the calibration solution with the
ChecktempC, and adjust the °C knob accordingly.
Turn the rotary knob to select the 19990 µS/cm range.
Turn the k% calibration knob until the display shows the
conductivity reading at 25°C (see "Conductivity vs. temperature
chart "), e.g. 12880 µS/cm (= 12.88 mS/cm), and all
subsequent measurements will be compensated to 25°C (77°F).
If you prefer to standardize the temperature compensation to
20°C (68°F), adjust the trimmer to read "11.67 mS" (see
"Conductivity vs. temperature chart "). All subsequent measure-
ments will be compensated to 20°C.
The calibration is now complete and the meter is ready for use.
PROCEDURE FOR HI 933000
Pour the conductivity calibration solution (e.g. HI 7030) into a
beaker while making sure that the holes on the probe are
completely submerged. If possible, use plastic beakers to minimize
any EMC interference.
HI 7030
°C
CALIBRATION
Each meter is supplied complete with a 9V battery. Slide off the
battery compartment cover on the back of the meter and install
the battery while paying attention to its polarity (see "Battery
Replacement" section for details).
Connect the probe to the meter by aligning the pins with the
socket and pushing the plug in. With HI 933000, tighten the
threaded ring.
Make sure that the meter has been calibrated before taking any
measurements (see "Calibration" section for details).
Immerse the conductivity probe into the
sample, making sure that the holes on
the shaft are completely submerged.
If possible, use plastic beakers to minimize
any EMC interference.
Tap the probe lightly on the bottom of the beaker to remove any
air bubbles which may be trapped inside the sleeve.
Turn the instrument on by pressing the ON/OFF key or by setting
the rotary switch (HI 8033) to the desired measurement range.
For HI 8033:
Measure the solution temperature with a
ChecktempC or another accurate ther-
mometer, and set the temperature knob
to the measured value (e.g. 20°C).
Select the appropriate measurement range.
Note: If the display shows only a "1", the
meter is out of range. Select the next
(higher) range.
For HI 933000:
Before taking any measurement, wait a couple of minutes for the
temperature sensor to reach thermal equilibrium with the sample.
When the temperature of the sample is lower than 20°C or higher
than 30°C, allow more time for the thermal equilibrium to be
achieved.
After measurements, switch the instrument off and clean the
probe (see "Probe Maintenance" section for details).
OPERATIONAL GUIDE
°C