Instruction Manual HI 8033 HI 933000 Portable Multi-Range Conductivity Meters w w w . h a n n a i n s t .
Dear Customer, Thank you for choosing a Hanna Product. Please read this instruction manual carefully before using the instrument. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com. These instruments are in compliance with the directives. WARRANTY All Hanna Instruments meters are warranted for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions.
FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 8033 FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 933000 mS 1 Conductivity meter HI 8033 ppm CaCO3 salinity 0-19,990 µS 0-199,9 µS 0-1999 µS 0-19,990 OFF 2 °C 0 k% 10 + 20 3 10 15 10 15 5 4 0 30 - 50 40 5 ins rum n s 1) 2) 3) 4) Liquid Crystal Display Rotary switch Knob for manual temperature compensation Calibration knob Range 0.0 to 199.9 ; 0 to 1999 µS/cm 0.00 to 19.99 mS/cm 0 to 19990 ppm Resolution 0.1 ; 1 µS/cm / 0.
OPERATIONAL GUIDE CALIBRATION • Each meter is supplied complete with a 9V battery. Slide off the battery compartment cover on the back of the meter and install the battery while paying attention to its polarity (see "Battery Replacement" section for details). • Connect the probe to the meter by aligning the pins with the socket and pushing the plug in. With HI 933000, tighten the threaded ring.
• Immerse the conductivity probe in the solution and wait a couple of minutes for thermal equilibrium to be reached. • Tap the probe on the bottom, then shake it while rotating to make sure no air bubbles remain trapped in the sleeve. • Switch the instrument on by pressing ON/OFF, and select the 19.99 mS/cm range by pressing the corresponding range key. • With the supplied screwdriver, adjust the calibration trimmer in the battery compartment until the display shows "12.88 mS", i.e.
PROBE MAINTENANCE After every series of measurements, rinse the probe with tap water. If a more thorough cleaning is required, remove the PVC sleeve and clean the probe with a cloth or a non-abrasive detergent. When reinserting the sleeve onto the probe, be sure that the sleeve is in the right direction with the holes towards the cable end. After cleaning the probe, re-calibrate the instrument. The probe body is made of PVC. For this reason it must never come into close contact with a heat source.
ACCESSORIES CALIBRATION SOLUTIONS CONDUCTIVITY PROBES HI 7030L HI 7030M HI 7031L HI 7031M HI 7033L HI 7033M HI 7034L HI 7034M HI 7035L HI 7035M HI 7039L HI 7039M HI 7032L HI 7032M HI 7036L HI 7036M 12880 µS/cm, 500 mL bottle 12880 µS/cm, 230 mL bottle 1413 µS/cm, 500 mL bottle 1413 µS/cm, 230 mL bottle 84 µS/cm, 500 mL bottle 84 µS/cm, 230 mL bottle 80000 µS/cm, 500 mL bottle 80000 µS/cm, 230 mL bottle 111800 µS/cm, 500 mL bottle 111800 µS/cm, 230 mL bottle 5000 µS/cm, 500 mL bottle 5000 µS/cm, 230 mL bo
CE DECLARATION OF CONFORMITY HI 8033 and HI 933000 Recommendations for Users Before using these products, make sure that they are entirely suitable for the environment in which they are used. Operation of these instruments in residential area could cause unacceptable interference to radio and TV equipment, requiring the operator to take all necessary steps to correct interferences. The metal rings of the probe are sensitive to electrostatic discharges. Avoid touching these rings at all times.
SALES AND TECHNICAL SERVICE CONTACTS Australia: Tel. (03) 9769.0666 • Fax (03) 9769.0699 China: Tel. (10) 88570068 • Fax (10) 88570060 Egypt: Tel. & Fax (02) 2758.683 Germany: Tel. (07851) 9129-0 • Fax (07851) 9129-99 Greece: Tel. (210) 823.5192 • Fax (210) 884.0210 Indonesia: Tel. (21) 4584.2941 • Fax (21) 4584.2942 Japan: Tel. (03) 3258.9565 • Fax (03) 3258.9567 Korea: Tel. (02) 2278.5147 • Fax (02) 2264.1729 Malaysia: Tel. (603) 5638.9940 • Fax (603) 5638.9829 Singapore: Tel. 6296.