User Guide

37
To obtain accurate readings, use the calibration solution in
the selected range and closer to the values to be measured.
Offset Calibration
To perform the EC or TDS calibra-
tion enter the calibration mode, by
pressing CAL and entering the pass-
word.
After the correct password is en-
tered, the control actions stop and
the primary LCD will display the first
EC or TDS calibration value, with
the "CAL" indicator blinking. The sec-
ondary LCD displays the
temperature.
Note If the wrong password is entered the system reverts back to
normal operation, displaying EC or TDS values.
0 is the default value for the 1
st
cali-
bration point. Dry the conductivity
probe and leave it in air.
Only when the reading is stable the
"CAL" indicator will stop flashing (af-
ter about 30 seconds) and the
"READY" and "CFM" indicators will
start blinking.
Press CFM to confirm the calibration
point; the primary LCD will display
the second expected buffer value.
If the zero calibration cannot be per-
formed, "ERROR" will blink.
Cell constant calibration
Select the solution value on the
primary display by pressing
or if the selected range has
two possibilities (e.g. 5.000 and
12.880 mS).
Immerse the EC/TDS probe with the temperature sensor in
the selected solution. The level of solution must be higher