If asking for last calibration data and the controller was never
calibrated, it answers with “0”; e.g. “01<STX>0<ETX>”.
If the controller was calibrated, it answers with “1” followed
by the calibration data.
The Data field of the answer has the following format:
1<Date><Time><CellCostant>
Date: DDMMYY (e.g. “170400” for April 17,2000)
Time: HHMM (e.g. “1623” for 4:23 pm)
Cell Costant: ASCII string (e.g. “1200”)
The items in the Data field are separated by blanks.
SETTING THE BAUD RATE
The transmission speed (baud rate) can be selected with the
up and down keys; then press the CFM key to store the new
settings.
Available baud rates are: 1200, 2400, 4800 and 9600 bps.
control action is active, no alarm condition is present and
controller setup is modified (must update controller setup for
PC - GET command for setup items).