User Guide

9
9. If everything is satisfactory the LCD will
display "Stor" and then the expected
value for the second buffer (two-point
calibration). If a wrong solution or elec-
trode has been used or if the buffer is
polluted, "WRONG" will be displayed to
alert the user.
10. If a single point calibration is required,
press (ALT+) CAL to exit the calibration
mode and maintain the previous slope
calibration. The instrument then checks
the electrode parameters and advises user
of possible abnormalities by "old probe"
and "dead probe" indications.
11. Press or to select the second buffer value. The meter will
display the "BUF 2" indication.
12. Rinse the electrode with some of the second buffer solution or clean
water.
13. Dip the tip (4 cm/1½") of the pH electrode into a beaker
containing the second buffer.
14. When the "CFM" symbol blinks, press CFM
to confirm the second calibration point.
15. The LCD will display "Stor". The instrument checks the electrode
parameters and advises the user of possible abnormalities by "old
probe" and "dead probe" indications (in these cases, repeat the
calibration procedure with fresh buffers). If everything is satisfactory
the meter is calibrated and it returns to normal operational mode.
Note: To abort the calibration procedure and leave the calibration data
unchanged, press ALT and CAL before confirming the first point.
CALIBRATION ERROR MESSAGES
If the "old probe" or "dead probe" messages are displayed, or if the
WRONG tag blinks during calibration, check the electrode by following
the conditioning and maintenance procedures and repeat calibration.
If calibration can not be successfully performed, the pH electrode
might have to be replaced.
Note: See "GLP" section for more details about "old probe" and "dead
probe" messages.