User Guide
8
CALIBRATION
pH CALIBRATION
For greatest accuracy, it is recommended to frequently calibrate the
instrument. For a faster procedure, it is possible to calibrate at 1
point, but it is always a good practice to calibrate at 2 points.
For a two-point calibration two of the following pH buffers can be
used: pH 4.01, pH 7.01 (or 6.86), pH 10.01 (or 9.18).
In the case of a two-point calibration in the acidic range (from 0 to
7 pH), use pH 7.01 (or 6.86) buffer as first solution and pH 4.01
buffer as second solution. If testing in the alkaline range (from 7 to
14 pH), use pH 10.01 (or 9.18) buffer as second solution.
Due to electrode conditioning time, the electrode must be kept
immersed a few seconds to stabilize. The meter is equipped with a
stability indicator and the user will be guided step by step with easy
indications on the LCD during the calibration. This will make the
calibration a simple and error-free procedure.
PROCEDURE
1. Rinse the electrode with a portion of the first calibration buffer or
clean water. Dip the tip (4 cm/1½") of the electrode into a
beaker containing the solution.
2. Press (ALT+) CAL when the meter dis-
plays pH measurement.
3. Enter the password (if different from "0000") with the arrow keys.
4. Press CFM to confirm the password or (ALT+) CAL to exit.
5. If password is correct, the meter displays "7.01pH" on the
secondary LCD with the "BUF 1" and "CAL" indications. The
primary LCD shows the uncalibrated pH reading.
Note: The pH buffer value (displayed on the secondary LCD) varies
with temperature. For example, at 20
°
C it shows 4.00 -7.03 -
10.06, at 25
°
C it shows 4.01-7.01-10.01.
6. Select the first buffer solution value with or if necessary.
7. When the "CFM" symbol blinks, the read-
ing is stable and calibration can be
confirmed.
8. Press CFM to confirm the first buffer.