User Guide

25
ACCESSORIES
pH CALIBRATION SOLUTIONS
HI 70004P pH 4.01 buffer, 20 mL sachet (25 pcs)
HI 70007P pH 7.01 buffer, 20 mL sachet (25 pcs)
HI 70010P pH 10.01 buffer, 20 mL sachet (25 pcs)
HI 7004L pH 4.01 buffer, 500 mL bottle
HI 7006L pH 6.86 buffer, 500 mL bottle
HI 7007L pH 7.01 buffer, 500 mL bottle
HI 7009L pH 9.18 buffer, 500 mL bottle
HI 7010L pH 10.01 buffer, 500 mL bottle
ELECTRODE STORAGE SOLUTION
HI 70300L Storage solution, 460 mL bottle
ELECTRODE CLEANING SOLUTIONS
HI 70000P Electrode Cleaning solution, 20 mL sachet (25 pcs)
HI 7061L General Cleaning solution, 460 mL bottle
HI 7073L Protein Cleaning solution, 460 mL bottle
HI 7074L Inorganic Cleaning solution, 460 mL bottle
HI 7077L Oil & Fat Cleaning solution, 460 mL bottle
REFILL ELECTROLYTE SOLUTIONS
HI 7071 3.5M KCl + AgCl Electrolyte, 50 mL (4 pcs)
HI 7072 1M KNO
3
Electrolyte, 50 mL (4 pcs)
HI 7082 3.5M KCl Electrolyte, for double junction electrodes, 50
mL (4 pcs)
Dry Membrane/Junction: Soak in Storage Solution HI 70300
for at least 1 hour.
Drifting: Soak the electrode tip in warm Hanna Solution HI
7082 for 1 hour, then flush tip with distilled water. Refill with
fresh HI 7071 for single junction electrodes and HI 7082 for
double junction electrodes.
Low Slope: Refer to the cleaning procedure above.
No Slope: Check the electrode for cracks in glass stem or bulb
and replace the electrode.
Slow Response/Excessive Drift: Soak the tip in Hanna Solu-
tion HI 7061 for 30 minutes, rinse thoroughly in distilled water
and then follow the Cleaning Procedure above.